-33-
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Guidance and manufacturer’s declaration - Emission
The [MDS3001/MDS3001PLUS/MDS3001U] is intended for use in the electromagnetic environment
specified below. The customer or the user of the [MDS3001/MDS3001PLUS/MDS3001U] should assure
that it is used in such an environment.
Phenomenon
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1, Class B
Home healthcare environment
Harmonic distortion
IEC 61000-3-2
Class A
Home healthcare environment
Voltage fluctuations and flicker
IEC 61000-3-3
Compliance
Home healthcare environment
For all ME EQUIPMENT and ME SYSTEMS