Troubleshooting Guide
Situation
Possible Cause
Solution
Free Flow (“FrE FLO”)
alarm activated*
Silicone tubing detached from pump rotor;
formula is gravity flowing
Assemble silicone tubing properly around pump rotor and
secure in pump adapter bracket and tubing guide per dia-
grams 3, 4, 5 on pages iv
Drip chamber walls coated with formula or condensation
in sensor path
Manipulate chamber to clear drops or condensation in
sensor path
Drip sensor is in direct path of intense light
Turn pump so sensor is not in direct sunlight
Flow sensors need cleaning
Clean with cotton swab dampened with isopropyl alcohol
DOSE COMPLETE
light is on and alert
sounds
Pump has delivered preset volume
Press RESET. Press RUN/PAUSE.
Pump will operate until dose is completed again
BATTERY CHARGING
light is on
AC Power is available and the battery is in charging mode
Note:
Bat Charging light indicator only available on serial
numbers greater than 850000
N/A
Audible Alert sounds
when , or RESET is
pressed
When pump is running it is designed to give audible alert
when one of these buttons is pressed
N/A
* For Model #199235 with serial numbers greater than 400000
** Model 199235 only
If the pump serial number is less than or equal to 400000 and the pump RUN/HOLD button is colored RED, refer to section titled Free Flow
Alarm Confirmation Test to determine if the Free Flow Alarm has been installed on the pump.