30
Click the button
【
Default
】
, the print settings will go back to the default.
Each item, its options and explanation are shown in the table below.
Item
Options
Explanation
Print Mode
[Auto 3×4+1]/[Auto 3×4]/
[Auto 2×6+1]/[Auto 2×6]/
[Auto 4×3]/[Auto 3-2]/
[Auto 3-2+1]/[Rhythm 4]/
[Rhythm 3 ]/[Rhythm 2]/[Manual]
The selection will be used as the default printing
mode.
Lead Gain
Smart/Current
The option selected will be used as print gain mode.
"Smart" means the system will adjust gain
automatically to fit paper height; "Current" means it
will use screen waveform gain as that of printing.
Auto Strip
3Sec/4Sec/5Sec
/6Sec/8Sec/10Sec/
15Sec/20Sec/25Sec
The selection will be used as the time for printing
each strip.
Rhythm Strip
10Sec/15Sec/20Sec
/25Sec/30Sec
In printing mode, when selecting rhythm 2, rhythm
3, or rhythm 4, the system will use the time selected
as the time for printing each row waveform.
Average QRS
[4×3+Mark]/[4×3]/
[3×4+Mark]/[3×4]/
[2×6+Mark]/[2×6]/[None]
In printing mode, when selecting "Rhythm" or
"Auto", the system will use the format selected to
print average QRS waveform.
Auto-Diag
All/Data/Conclusion/None
Printing diagnosis includes data and conclusion,
user can select as his need.
Periodic
[per1Min]/[per 2 Min]
/[per3Min]/[per5Min]
/[per10Min]/[per20Min]
/[per30Min]/
[per 60 Min]/[Off]
When gathering ECG, the system will activate
printing according to the duration selected. When
printing mode is manual, it will print in "3 x 4+1"
mode, or in the mode currently set.
Note: "Auto Strip", "Rhythm Strip", "Average QRS", "Auto-Diag", "Periodic" are only
available in auto and rhythm print mode.
13.12 Checking Electrodes Placement
Select the
button in the main interface to view the sketch map of electrodes placement
shown below.