3. Operation  
 
Assembly of the instrument

 

 
1. 

Place  thermometer  support  with  glass  thermometer  onto  the  rim  of  the  bath 
opposite the control panel, 

making sure that the thermometer does not touch 

the floor of the bath. 

 

 

 
 
 
 
 
 

 

(The  fixing  and  height  adjustment  of  the  thermometer  on  the  thermometer 
support  is done by the enclosed rubber ring being rolled over the thermometer 
from above) 
 

 

 

2. 

Fill  with  distilled  or  fully  desalinated  water  up  to  1-2  cm  below  the  rim  of  the 
bath.  If  tap  water  is  used,  the  black  interior  surface  will  become  white  due  to 
lime deposits. The sections are then not clearly visible any more. 
 

3. 

Before  connecting  the  instrument  with  the  enclosed  line  cord  to  the  grounded 
current  supply,  it  is  necessary  to  check  that  the  power  supply  is  the  same  like 
prescribed on the type plate of the instrument. 
 

4. 

Connect unit to mains power. 
 

Summary of Contents for TFB 55

Page 1: ...Read this manual before starting the instrument Rev 00 04 2010 Copyright Duplication of this manual by any means for any purpose without written consent of Medite GmbH is prohibited MEDITE GmbH Wollen...

Page 2: ...E Guidelines indicated below including all changes and addendums to date thereto The above mentioned product has been controlled and meets the following standards and guidelines EU Directive 2004 108...

Page 3: ...e 1 1 3 Authorized Operator 1 1 4 Safety measures on site 1 2 Transport Installation 2 2 1 Technical data 2 2 2 Operation conditions 2 2 3 Transport Storage 3 2 4 Unpacking 3 2 5 Set up and connection...

Page 4: ...reasons For the exchange of defective parts only original medite spare parts have to be used The conditions for operation maintenance and service mentioned in this instruction manual have to be strict...

Page 5: ...system avoids overheating and damage even in case of empty bath Equipment mains switch temperature control with scale in C 2 2 Operating conditions The operation is only allowed in closed rooms labor...

Page 6: ...whether the equipment is complete The TFB 55 is supplied with one power cord and one instruction manual 2 5 Installation and Connection Before starting the instrument please make sure that your mains...

Page 7: ...ne by the enclosed rubber ring being rolled over the thermometer from above 2 Fill with distilled or fully desalinated water up to 1 2 cm below the rim of the bath If tap water is used the black inter...

Page 8: ...y turning the temperature regulator knob Always turn clockwise Pilot lamp lights up and extinguishes when the temperature is achieved Before the required temperature is achieved heating will be switch...

Page 9: ...damage the black epoxy coating In case of a non functioning of the instrument e g caused by voltage oscillations please check whether the fuses in the back of the instrument are defect Should the def...

Page 10: ...riod of use please deliver the equipment to an authorized disposal facility The company Medite as manufacturer is also prepared to take back the equipment for disposal against payment for the disposal...

Page 11: ...Rev 00 04 2010 HE TFB55e_Druck_00 DOC 8 7 Wiring diagram...

Page 12: ...ES 99 430 Width 360 mm Depth 570 mm Height 300 mm Cooling plate 210 x 260 mm Temperature adjustable from 15 C to 10 C Weight 20 kg Voltage 230V 50Hz 150VA Dispenser Unit TES 99 200 Width 360 mm Depth...