3 Operating

GB

19

WARNING

Clean and disinfect the nebulizer kit as described under "

4.1 

Cleaning and maintenance

" in order to avoid a possible infection

by contaminated parts!
Additionally general hygiene regulations (e.g. thoroughly washing
the hands, hygienic handling of the medication) must be observed.

3.1
Preparing
for
inhalation

NOTE

Before starting therapy, discuss the length, dosage and frequency 
of use with your doctor. The method of use should be appropriate 
to the aim of the therapy (e.g. lung treatment = use of mouthpiece
and treatment of the nasal-pharyngeal area = use of mask). 
Consult your doctor about this aspect as well.

w

r

t

t

w

e

w

1

u

  3.

  4.

5a.

5b. 

Fill  in  the  correct  amount  (min.  2  ml  and  max.  10  ml)  of  prescribed 
medication into the nebulizer tank      . If you have overfilled the tank 
pour  all  the  liquid  out  and  clean  the  nebulizer  kit  as  described 
under point "

4.1 Cleaning and maintenance

".                                        

g

Rotate the inhalation top      clockwise to properly fix it on the nebulizer
tank      .                                                                                                 

g

Plug either the mouthpiece      , or a nose adaptor (      or      ) with the 
connection tube for mouthpiece and nose adaptor     onto the inhalation 
top. Therefore the connection tube      needs to be put onto the nozzle 
on  the  upper  part  of  the  inhalation  top  with  the  longer  end.  Then  the 
mouthpiece       or the nose adaptor (      or      ) needs to be connected 
to the shorter end of the connection tube      .              

                          g

Alternatively, you may use the adult or child mask (      or      ) for the
inhalation. Therefore you need the mask connection tube instead of the 
connection tube      . Make sure, the power switch      is turned off ("O").
Connect  the  power  plug  to  the  mains  outlet  and  place  one  end  of  the 
air tube       firmly over the connection for air tube        of the compres-
sor. Place the other end over the appropriate connection for air tube       

of the nebulizer kit.

s

p

a

z

z

z

i

o

z

3

s

p

a

q

  1.

  2.

Rotate  the  inhalation  top              counterclockwise  to  remove  it  from  the
nebulizer tank     .                                                                                   

g

Place the vaporizer head      on the nozzle      .

NOTE

The are two additional vaporizer heads included in the scope of 
delivery of model IN 550 Pro (art.no. 54530). You have the possi-
bility to use one of totally 3 heads (blue     , red     or yellow     ) on
the nozzle      . This allows a more targeted application. Particle
sizes of approx. 4.5 to 6.0 µm (red vaporizer head) are ideal for use
on mouth and throat (e.g. laryngitis). Particle sizes of approx. 3.5
to 4.5 µm (blue vaporizer head) are ideal for use on the respiratory 
tract and lungs front (e.g. bronchitis) and smaller particle sizes of 
approx. 3.0 to 4.0 µm (yellow vaporizer head) are ideal for the 
deeper areas of the lung (e.g. asthma). Consult your physician
to find out, which particle size is ideal for your personal situation.

r

g

h

e

r

/    :

g

   :

h

Summary of Contents for 54520

Page 1: ...uchsanweisung Manual Kullan m talimat Bitte sorgf ltig lesen Please read carefully L tfen dikkatle okuyunuz 0297 DE GB TR Inhalator IN 500 Compact IN 550 Pro Inhalator IN 500 Compact IN 550 Pro Neb li...

Page 2: ...es 3 Anwendung 4 Verschiedenes 5 Garantie 1 5 6 8 13 1 Safety Information 2 Useful Information 3 Operating 4 Miscellaneous 5 Warranty 14 18 19 21 26 TR Kullan m talimat 1 G venlik bilgileri 2 Bilinmes...

Page 3: ...Ger t und Bedienelemente Device and controls Cihaz ve Kullanma Elemanlar 1 2 3 6 7 0 u i o p a s d f t w q e r z 4 8 9 Art 54530 Art 54520 5 7 8 nur only sadece Art 54530 g h...

Page 4: ...size 4 5 6 0 m only art no 54530 Vaporizer head yellow particle size 3 0 4 0 m only art no 54530 DE Kompressoreinheit 1 2 3 4 5 6 7 8 9 0 q w e r t z u i o p a s d f g h Verneblerset Ger te Anschluss...

Page 5: ...apa Hava hortumu i in neb liz r ba lant s Neb liz r tank A z P sk rtme ba l mavi partik l b y kl 3 5 4 5 m Neb liz r st par as A zl k ve burun adapt r i in ba lant par as Hava hortumu Maske ba lant pa...

Page 6: ......

Page 7: ...en zur Inbetriebnahme und Handhabung Lesen Sie diese Gebrauchsanweisung voll st ndig Die Nichtbeachtung dieser Anweisung kann zu schweren Verletzungen oder Sch den am Ger t f hren WARNUNG Diese Warnhi...

Page 8: ...utzt zu werden es sei denn sie werden durch eine f r ihre Sicher heit zust ndige Person beaufsichtigt oder erhielten von ihr Anweisungen wie das Ger t zu benutzen ist g Kinder m ssen beaufsichtigt wer...

Page 9: ...nd verwenden Sie es nicht in der N he von Hitzequellen g Um die Gefahr eines Stromschlags zu vermeiden ziehen Sie sofort den Netzstecker aus der Steckdose g bei St rungen w hrend der Benutzung g nach...

Page 10: ...werden g Kontrollieren Sie vor jeder Anwendung das Geh use das Netzkabel den Netzstecker den Luftschlauch und die Schlauchanschluss ffnungen und auf Unversehrtheit und pr fen Sie ob das Ger t normal f...

Page 11: ...fen Sie zun chst ob der Lieferumfang vollst ndig ist und der Artikel keinerlei Besch digungen aufweist Im Zweifelsfalle benutzen Sie den Artikel nicht und kontaktieren eine Servicestelle G Zum Liefer...

Page 12: ...0 m roter Zer st uberkopf eignen sich insbesondere f r die Anwendung im Mund und Rachenraum z B bei Kehlkopfentz ndungen Partikel gr en von ca 3 5 bis 4 5 m blauer Zerst uberkopf sind besser geeignet...

Page 13: ...e Anwen dung wird besonders von Kindern als angenehmer empfunden o i p a s Maske Halten Sie die Maske leicht ber die Nase gepresst Die Maske muss die Nase dicht umschlie en jedoch ohne zu dr cken g Na...

Page 14: ...ausatmen Inhalieren Sie nicht zu schnell Wenn Sie eine Pause machen m chten unterbrechen Sie kurz WARNUNG Eine Inhalationsanwendung sollte 20 Minuten nicht ber schreiten Wird der Kompressor l nger bet...

Page 15: ...Tage muss der Luftfilter im Kompressor gewechselt werden Hierzu gehen Sie wie folgt vor G 2 Entfernen Sie die Abdeckung des Luftfilters in Pfeilrichtung Entfernen Sie den alten Filter z B mit einem ne...

Page 16: ...rst uberkopf fehlt oder keine bzw zu wenig Inhalations l sung im Vernebler tank Setzen Sie das Verneblerset korrekt zusammen F llen Sie min 2 und max 10 ml Inhalationsl sung in den Verneblertank Keine...

Page 17: ...aten entsprechen Sollte es not wendig werden die Sicherungen auszutauschen verfahren Sie wie folgt 1 2 3 ffnen Sie den Deckel f r das Netzkabelfach auf der R ckseite des Kom pressors g Durch gleichzei...

Page 18: ...4 5 m Zerst uberkopf blau 0 2 ml Min 65 dB 205 kPa 100 190 kPa 10 C 40 C rel Luftfeuchte 85 10 C 40 C rel Luftfeuchte 95 1 3 kg Art 54520 2 0 kg Art 54530 19 5 x 13 5 x 9 2 cm Art 54520 33 x 20 x 13...

Page 19: ...uteile ein 4 Von der Garantie ausgeschlossen sind a alle Sch den die durch unsachgem e Behandlung z B durch Nichtbeachtung der Gebrauchsanweisung entstanden sind b Sch den die auf Instandsetzung oder...

Page 20: ...served to prevent any damage to the device NOTE These notes give you useful additional information on the installation or operation Lot number Serial number Manufacturer Explanation of symbols Date of...

Page 21: ...rding to the instructions the user is liable for the safe operation of the unit g A single inhalation application should not exceed 20 minutes g Only use genuine accessories and spare parts provided b...

Page 22: ...d over sharp edges or allow it to be pinched do not run the appliance nearby heat sources g To avoid the risk of electric shock immediately pull out the plug from the power source g if a malfunction o...

Page 23: ...use multi sockets or extension cables g Do not fill more than 10 ml of a suitable inhalation liquid into the nebuli zer tank g Before using the device make sure that all parts are correctly assembled...

Page 24: ...NA Inhalator unit we recommend that you read the following operating and maintenance instructions carefully Please check first of all that the unit is complete and is not damaged in any way If in doub...

Page 25: ...h the longer end Then the mouthpiece or the nose adaptor or needs to be connected to the shorter end of the connection tube g Alternatively you may use the adult or child mask or for the inhalation Th...

Page 26: ...e mouthpiece out of your mouth Then put it into your mouth again lips enclosing it tightly and breathe slowly in and out again s 3 Alternatively a child inhalation mask an adult inhalation mask g and...

Page 27: ...ox 40 C without any detergent again The nebulizer parts now need to be disinfected please refer to the chapter Disinfection G Clean the compressor case with a soft damp cloth Do not use any aggressive...

Page 28: ...t clean the air tube with cleaning or disinfecting solutions If the outside of the air tube is dirty wipe it clean with a soft cloth moistened with water or mild detergent If the inside of the air tub...

Page 29: ...esp too less medication in the nebulizer tank Assemble the nebulizer kit in the correct way Fill in medication min 2 ml max 10 ml into the nebulizer tank No nebulization or low nebulization rate The n...

Page 30: ...ing to the specifications mentioned in chapter 4 8 Technical specifications If it is necessary to replace the fuses follow the below steps 1 2 3 Open the mains cable com partment lid on the rear side...

Page 31: ...ml 5 m 0 2 ml min 65 dB 205 kPa 100 190 kPa 10 C 40 C rel humidity 85 10 C 40 C rel humidity 95 1 3 kg Art 54520 2 0 kg Art 54530 19 5 x 13 5 x 9 2 cm Art 54520 33 x 20 x 13 8 cm Art 54530 54520 5453...

Page 32: ...or invoice 2 Defects in material or workmanship will be removed free of charge within the warranty period 3 Repairs under warranty do not extend the warranty period either for the unit or for the rep...

Page 33: ...nu da mutlaka birlikte verin Bu kullan m k lavuzu bu cihaza aittir Devreye alma ve kullan m ile ilgili nemli bilgiler i erir Bu kullan m k lavuzunu ba tan sona okuyunuz Bu k lavuza uyulmamas a r yaral...

Page 34: ...uygulaman n s resi 20 dakikadan uzun olmamal d r i Sadece retici firmaya ait orijinal aksesuarlar ve yedek par alar kullan n z i Sadece solunum tedavisi i in uygun s v halindeki ila lar kullan n Ciha...

Page 35: ...k t rmay n ve s kaynaklar n n yak n nda kullanmay n i Elektrik arpmas tehlikesini nlemek i in a a daki durumlarda elektrik fi ini derhal prizden ekin i Kullan m s ras nda ar zalarda i her kullan mdan...

Page 36: ...e veya bir ar zan n bulundu unu tahmin ediyorsan z neb lizat r al t r lmamal d r Derhal elektrik fi ini prizden ekin ve servis noktas yla ileti ime ge in i Cihaz elektrik beslemenize ba lamadan nce ti...

Page 37: ...kullan m ve bak m uyar lar n dikkatle okuman z nermekteyiz L tfen nce cihaz n eksik veya hasarl olup olmad n kontrol ediniz Emin de ilseniz cihaz al t rmay n ve bir servis yerine g nderin Teslimat ka...

Page 38: ...ya burun adapt r veya ba lant par as n n k sa taraf na i itilir Alternatif olarak neb lizasyon i in yeti kin veya ocuk maskesini ve ya kullanabilirsiniz Bunun i in ba lant par as n n yerine maske ba l...

Page 39: ...n a z n zdan kartabilirsiniz Tekrar a z n za al p dudaklar n z s k ca kapat n z yava a nefes al p veriniz b Maske ya da burun adapt r ile neb lizasyon i A zl a alternatif olarak teslimat kapsam nda bi...

Page 40: ...rulay n Neb liz r par alar n n sadece dezenfekte edilmeleri kal r Dezenfeksiyon k sm na bak n z i Kompres r yumu ak hafif nemlendirilmi bir bezle silebilirsiniz Temiz leme i in agresif maddeler rn ovm...

Page 41: ...kalan nem nedeniyle bakteri olu umunu engellersiniz i Hortumu temizleyiciler veya ba ka dezenfeksiyon y ntemleri ile temiz lemeyin Hortumun d y zeyini hafif sabun zeltisine dald r lm bir bezle silebil...

Page 42: ...eya neb liz r tank nda neb l zeltisi yok ya da yetersiz Neb liz r setini do ru bir ekilde birle tirin Neb liz r tank na en az 2 ve en fazla 10 ml neb l zeltisi doldurun Neb lizasyon g c yok veya ok az...

Page 43: ...rar kapat n IN 550 Pro kompres r n arkadan g r n m 9 Bu neb lizat r cihaz 93 42 EWG Y netmeli i nin Medikal r nlere ili kin AB Y netmeli i ve EN 13544 1 2007 Solunum tedavi cihazlar B l m 1 Neb liz r...

Page 44: ...l dak 10 ml 5 m 0 2 ml dak 65 dB 205 kPa 100 190 kPa 10 C 40 C ba l hava nemi 85 10 C 40 C ba l hava nemi 95 1 3 kg r n 54520 2 0 kg r n 54530 19 5 x 13 5 x 9 2 cm r n 54520 33 x 20 x 13 8 cm r n 545...

Page 45: ...umlar a Uygunsuz kullan m rne in kullan m talimat na dikkat edilmemesi sebebiyle olu mu olan t m zararlar b Al c n n veya yetkisiz nc ah slar n onar m na veya m dahalesine dayand r labilecek zararlar...

Page 46: ...40 TR...

Page 47: ...CO LTD No 8 Tongxing Road Economic Technological Development Area 226009 Nantong City PEOPLE S REPUBLIC OF CHINA thalat Firma KALE ELEKTRON K DI T CARET A BAKIR VE PIRINCCILER SAN SIT MENEKSE CAD NO 2...

Page 48: ...MEDISANA AG Jagenbergstra e 19 41468 NEUSS Deutschland eMail info medisana de Internet www medisana de 54520 30degbtr 06 2017 Ver 1 1...

Reviews: