Disconnect the power supply unit from the mains during thunderstorms or if the device is not used
for a longer period of time.
3.4.
Power Supply Connection
medinSINDI
®
is immediately ready for use if the battery is charged. If the battery is depleted, the power supply
unit must be connected.
3.4.1.
Initial start-up and long-term storage:
For the initial start-up and after a longer period of storage, the external power adapter provided must be
connected in order to charge the battery. The charging time is approx. 5 hours. An oxygen and pressure
calibration must be performed before the initial start-up (see 4.4.3.
3.4.2.
Connection
Plug external power adapter into the connector provided (see Figure 3)
3.4.3.
Operating and Idle Times
-
Without external power supply:
Operation for 5 hours is possible with a fully charged battery for brief operating or transport times.
-
With external power supply: Continuous operation possible (ICU)
Important Note – Battery
When the battery is nearly depleted, an alarm sounds once per minute for 1 second.
The charging process is shown on the medinSINDI
®
monitor when the device is switched off.
Figure 2: Symbol to indicate the charging process
3.4.4.
Interruption of the External Power Supply
The power supply can be disconnected or connected during operation without any effect on functionality or
loss of data.
3.4.5.
Background Illumination Display
-
brightly lit when external power supply is used
-
dimmed when internal battery is used (energy-saving mode)
-
off when battery is nearly depleted
3.4.6.
Power supply connection:
To disconnect the medinSINDI
®
from the mains, the plug of the power supply unit of the medinSINDI
®
must be
pulled out of the electrical outlet. Otherwise the medinSINDI
®
is still connected to the mains, even when it is
turned off.
Therefore, when setting up the medinSINDI
®
, it should be ensured that the plug of the power supply unit is still
accessible even when set up and the medinSINDI
®
can thus be easily disconnected from the mains.
Disconnect the power supply unit from the mains during thunderstorms or if the device is not used for a longer
period of time.