The prescriber or caregiver has the responsibility to be aware of all the available
adjustments, and ensure the product is correctly set up to meet the user’s needs.
Medifab recommends that a formal written record is maintained for all of the
setups and maintenance checks. A product log form is available at the end of
this manual.
Before placing the user in the seat, ensure the following adjustments are made.
If the caregiver has any questions or if additional information is required,
contact the user’s health care provider or Medifab.
Seat Depth
Remove or push aside the seat padding to reveal the four screws
(A)
located on
the seat base.
Undo the four screws
(A)
by a few turns with the hex allen key. Once the seat
depth bracket
(B)
is free, position the bracket to achieve the desired seat depth.
Secure the four screws tightly.
1. Seat depth
2. Seat width
3. Backrest height
4. Footplate height
Tools required to adjust your Shuttle Discovery are a 10mm ring spanner/socket
and 4mm hex allen key.
NOTE:
Ensure the assembly is tight. Do not adjust the seat
depth while the user is seated in the seat.
Setting up the Discovery Seat
A
A