All the control solution test results are record with CONTROL SOLUTION ICON
‘ ’ to distinguish them from those of actual blood glucose tests in the meter
memory. Recorded control solution test results will not be calculated in your
result averages.
If your control solution test results continue to fall outside the range printed
on the test strip vial, the GlucoDr.S Blood Glucose Monitoring System may
not be working properly.
- If this is the case, do not use the system to test your blood glucose levels.
- Contact your local distributor.
Out-of-range test results may be caused by one or more of the followings:
- Expired or contaminated control solution,
- Expired or damaged test strip,
- Use of control solution or test strip past its discarding date,
- Error in performing the test,
- Failure to shake the control solution vial well,
- Meter, test strips or control solution are kept too warm or too cold,
- or Meter malfunction