First installation / General preparations for safe operation and performance

Secure connection of your image source

In case, video peripherals are used within the medical video chain, it must be ensured that the image source is always connected to 

the live viewing monitor via a primary, direct connection. 
This primary, directly connected main monitor should always be used for live view. 
The medical video chain that is connected through the MVC Pro SDI to HDMI convertor can only be used for recording and extra 

viewing, e.g. educational purpose. It can’t be used for medical purpose.
If the output signal of the peripheral device is used for live view [3], it is necessary to switch the monitor immediately to the primary 

signal [1] in case of failure or accidental disconnection of the peripheral device.

1]  Primary connection camera to monitor (any signal type)
2]  Quad-SDI connection to MVC Pro
3]  Quad-SDI loop out to monitor (optional)
4]  HDMI connection to recorder

Checks before you operate the device: 

The operation of this device does not require special maintenance, calibration, special hygienic precautions or the use of 

consumable components.

1. Turn on the unit from the back panel power switch.
2. Turn on your camera system(s) after MVC Pro has started its boot sequence.

Compatibility to other devices

The compatibility list for medical video sources can be found in 

“Appendix A: MVC Pro Specifications”

 in this document. 

MAIN MONITOR

CAMERA CCU

MVC Pro

Summary of Contents for MVC Pro SDI to HDMI

Page 1: ...C Pro SDI to HDMI Medical Video Converter User Guide English Applies to firmware 171214 and higher MVCPro UG01 EN 211004 OB For more translations available in pdf please visit www medicapture com support ...

Page 2: ...e which if not avoided may result in minor or moderate injury and or appreciable material financial and environmental damage This image indicates the manufacturer and appears next to the manufacturer s name and address This symbol informs the user that this user manual and translated copies can be downloaded from the MediCapture website at www medicapture com support This symbol indicates the equi...

Page 3: ... with UL EN IEC 60601 1 shall be kept outside the patient environment as defined on the systems standard Note To the user and or patient Any serious incident related to the product shall be reported to the manufacturer and to the competent authority of the Member State where the user patient is established CAUTION To reduce the risk of fire or shock hazard and annoying interference use the recomme...

Page 4: ...ications directly related to the product are presently unknown On the basis of the patient s general condition the doctor in charge must decide whether the planned use is possible or not For further notes and instructions please refer to the latest medical literature General functional description MVC Pro SDI to HDMI is a medical grade Ultra High Definition Video Converter The device does not feat...

Page 5: ...fied by the manufacturer 13 Use only with a cart stand tripod bracket or table specified by the manufacturer or sold with the apparatus When a cart is used use caution when moving the cart apparatus combination to prevent the cart from tipping over and to avoid injury 14 Unplug this apparatus during lightning storms or when unused for long periods of time 15 Disconnection Method Switch off AC powe...

Page 6: ...Use supplied power cable to connect 100V 240V outlet 4K HD out 2 Loop out SDI in A B C D A B C D 4K HD out 1 SDI loop output of the original SDI input SDI Input Various input options are possible for 4K SDI 4K or HD HDMI Output 2 l On connection to power supply mains O Off no connection to power supply mains Power Switch This MVC Pro connects to an SDI video source for example an endoscopic camera...

Page 7: ...signal of the peripheral device is used for live view 3 it is necessary to switch the monitor immediately to the primary signal 1 in case of failure or accidental disconnection of the peripheral device 1 Primary connection camera to monitor any signal type 2 Quad SDI connection to MVC Pro 3 Quad SDI loop out to monitor optional 4 HDMI connection to recorder Checks before you operate the device The...

Page 8: ...C Pro SDI to HDMI 4x SDI cable HDMI cable Y split power cable Rubber Feet 4 in 1 and User Guide Compliance and Certification EN 60601 1 2006 A1 2013 A12 2014 EN 60601 1 2 2015 EN 60601 1 6 2010 A1 2015 EN 50581 2012 Classification Class I Equipment Mode of operation continuous No AP APG No Applied Part Conforms to Medical Devices Directive 93 42 EEC Medical General medical equipment as to electric...

Reviews: