3
1.
Please use No. 7 manganese or alkaline dry cell (2 batteries); don’t not use other kinds of
batteries, otherwise it may cause machinery breakdown.
2.
Please don’t ensure the positive and negative poles of the battery are in right position; take out
the battery immediately after the battery has run down, and replace it with a new one. Otherwise,
it will cause heat, liquid leakage and fracture, causing damage to the machine.
3.
Please don’t maintain or repair the infrared thermometer by yourself. If there is a quality
problem with the product or if you have any questions about the correctness of measurement
thermometer, please contact your dealer or manufacturer.
4.
After the replacement of components (such as batteries) and the thermometer reaches the end
of its equipment life, please handle it according to local laws and regulations.
Technical Parameters
Product Name
Medical Infrared Thermometer
Model
and
Specification
XL-F02
Measurement Range 32.0 ℃ ~ 42.9 ℃
Maximum
Permissible Error
The maximum allowable error of thermometer is ± 0.2 ℃ within 35.0 ℃
~ 42.0 ℃; the maximum allowable error is ± 0.3 ℃ beyond 35.0 ℃ ~
42.0 ℃.
Display Resolution
0.1℃/℉
Measuring Distance
5cm
~
10cm
Operating
Environment
Temperature: 16 ℃ ~ 35 ℃
Relative Humidity: ≤ 80%
Atmospheric Pressure: 86kpa ~ 106kpa
Transportation and
Storage conditions
Temperature: - 20 ℃ ~ + 55 ℃
Relative Humidity: ≤ 93%
Atmospheric Pressure: 50kPa ~ 106kpa
Clinical
Repeatability
±0.3℃
Battery
d.c.3V (two AAA batteries)
Measurement (MM) 155mm × 115mm × 45mm (L x W x H)
Prompt Function
When the operating range of ambient temperature exceeds 16 ℃ ~ 35 ℃,
the thermometer shall prompt “Err” or stop providing temperature value.
When it’s on body temperature mode, if the measured temperature is
lower than 32.0 ℃, it will display “Lo ℃”; if it is higher than 42.9 ℃, it
will display the prompt signal of “Hi ℃”.
Safety Requirements 1.1 Classified by type of anti-shock: equipment powered by internal
power supply.
1.2 Classified by the degree of anti-electric shock: B application part.
1.3 Classified by the protection degree of harmful liquid inlet: general
equipment.
1.4 Classified by safety degree when using flammable anesthetic gas