4134.A
TEMP 1
FLUID WARMER
OPERATORS MANUAL
MODEL 986S
Medical Solutions, Inc.
3901 Centerview Drive, Suite L
Chantilly, Virginia 20151
800-435-7600
703-834-0039 (Fax)
Page 1: ...4134 A TEMP 1 FLUID WARMER OPERATORS MANUAL MODEL 986S Medical Solutions Inc 3901 Centerview Drive Suite L Chantilly Virginia 20151 800 435 7600 703 834 0039 Fax ...
Page 2: ... will indicate the temperature of the top shelf heater plate The bottom display number in green will indicate the temperature of the bottom shelf heater plate During the time that the controller is directing the unit to warm a light will illuminate in the upper left corner of the display The Documentor is designed to assist in the compliance with JCAHO standard MM 2 20 and the fluid manufacturer s...
Page 3: ... Range Ambient to 43 C Factory Preset Temperature 40 C Ambient Operating Conditions Storage Temperature 40 to 80 C 40 to 176 F Humidity Conditions 0 to 90 up to 40 C non condensing Operating Temperature 10 to 55 C 14 to 131 F Atmospheric Pressure During minimum of 16 288 in Hg Transporting Electrical Requirements Power Requirements 110 120 VAC 50 60 HZ Fusing 2 each 1 5 Amps 2 each 0 5 Amps for th...
Page 4: ...r to the mounting bracket using these 3 screws 3 Plug the Documentor into the Hospital Grade Receptacle located on the side of the TEMP 1 4 Use the Mounting Diagram to place four mounting bolts either on a wall or a cart The TEMP 1 loaded with 6 bags of solution will weigh approximately 29 pounds When mounting to a wall care should be taken to secure the mount to the wall using bolts suitable for ...
Page 5: ...4134 A 5 Mounting Diagram TEMP 1 Time Clock Mounting Bracket Mounting Brackets ...
Page 6: ...eate a future date label See the Documentor Resetting Instructions for adjusting the clock date i Hold the Label sheet face up ii Line up the arrow on the Documentor under the words Remove Fluid By on the label iii Push the label sheet towards the arrow until the Documentor is activated iv Place a label on each IV bag for documentation and compliance purposes 2 Load the labeled bags of solution in...
Page 7: ...elow TO CHANGE THE SET POINT PRESS THE INDEX KEY LEFT MOST KEY TO VIEW SET POINT ONE SP1 TOP SHELF PRESS THE INDEX KEY TWICE LEFT MOST KEY TO VIEW SET POINT TWO SP2 BOTTOM SHELF PRESS THE UP OR DOWN ARROW KEYS TO CHANGE THE SET POINT PRESS THE ENTER KEY RIGHT MOST KEY TO ACCEPT NEW SET POINT PRESS THE INDEX KEY 4 The shelves will operate independently from each other When one shelf is not in use d...
Page 8: ...t button the small digits on the left representing the year will flash 13 Tap the left button until the desired number appears then Tap the right button The digits representing the month should flash 14 Tap the left button until the desired number appears then Tap the right button The digits representing the day should flash 15 Tap the left button until the desired number appears then Tap the righ...
Page 9: ...connect power to the unit and refer to qualified service personnel Remove the service panel qualified person and remove any liquid and or dry components Perform any necessary hospital electrical safety checks before returning the unit to service Trouble Shooting 1 If the unit fails to warm check the following a Check to see that the unit is plugged in and the power switch is on b Check the set poi...
Page 10: ...4134 A 10 Electrical Schematic ...
Page 11: ... the allotted time compare the temperature reading on the controller to the temperature reading on the Fluke Thermometer the two temperatures should be within one degree Celsius of each other If the two temperature readings are within one degree Celsius of each other the temperature controller has been verified If they are not please contact Medical Solutions Inc at 800 435 7600 Repeat the process...
Page 12: ...mptly notify MSI of any claim hereunder As a condition to the Warranty Buyer shall use and maintain the Equipment according to the provisions of the Operator s service manual IMPORTANT THIS IS YOUR TEMP 1 WARRANTY REGISTRATION Please register your warranty by filling out this page and send it to Medical Solutions Inc within 30 days of purchase A Hospital Name ______________________________________...