BATTERY BACK-UP 

SUPPLEMENT 

 

Summary of Contents for 061413B

Page 1: ...BATTERY BACK UP SUPPLEMENT ...

Page 2: ...ion against electrical shock 6 2 Class I 2 Protection against harmful ingress of water 6 3 None 3 Degree of safety in the presence of flammable anesthetics or oxygen 3 11 3 12 11 4 11 5 Not suitable for use in the presence of flammable anesthetics or oxygen 4 Mode of operation 6 6 Continuous Electromagnetic compatibility for immunity And emissions in accordance with IEC EN 60601 1 2 2007 ed 3 0 wi...

Page 3: ...000 12 Ceiling Wall Mount Battery Backup for System Two Light System 15 Section 4 Operation Basic Operation 16 Connecting the Battery 16 System Indicators 17 Indicators 17 Positioning the Unit 18 Section 5 Maintenance Fuse Replacement 19 Fuse Replacement Light Head 20 Fusing Specifications 20 Power Cord Replacement 20 Battery Replacement 21 Section 6 Troubleshooting General Troubleshooting 23 Tabl...

Page 4: ...uctor or both of the circuit at the service equipment or at the source of a separately derived system Neutral Conductor common In an AC circuit the return line for current Means of Isolation The Means of Isolation disconnects Isolates the Battery Back Up unit from the source of power This isolation makes it safe to work on the Battery Back Up or the light For the Floor mounted Battery Back Up the ...

Page 5: ...fety Instructions Installation Assembly shall be by qualified personnel only Installation of the ceiling wall unit shall be by a licensed electrician Maintenance section 5 shall be by a qualified person Follow all warnings and notices in Section 3 The power source for a ceiling wall mounted unit shall be on a circuit breaker connected to all energized supply conductors The fuses protect the circui...

Page 6: ... 1 Contact Medical Illumination via phone at 818 838 3025 through our website at http www medillum com or by facsimile at 818 838 3725 A Be prepared to give the model number serial number and full description of the failure B Our Customer Service department will attempt to solve the problem over the phone If it becomes necessary to send the product to the factory for repair you will be provided wi...

Page 7: ...maintenance or other repair services Damage to the product resulting from tampering accident abuse negligence alteration or other causes unrelated to problems with material and or workmanship Damage due to improper installation use cleaning or maintenance as outlined in the Installation and Service Manual for the product Labor costs associated with removing re packaging for shipment or reinstallin...

Page 8: ... 7 H x 19 W x 10 D Total Floor Height 77 H Electrical Specifications Parameters Value Input Voltage 100 240 VAC 50 60 Hz Single Phase Approximately 180W max Output Voltage 20VDC 26VDC Switch over time Approximately 5mS Back Up Power minutes MI 750 MI 1000 System Two LED At Full Intensity Minimum 300 minutes Minimum 240 minutes Minimum 200 minutes Environmental Specifications Parameter Value Ambien...

Page 9: ... for the MI 750 and the MI 1000 lights only The Ceiling Wall Mounted Battery Back Up can be used with the MI 750 MI 1000 and System Two LED ceiling mounted lights WARNING To avoid the risk of electric shock this equipment must only be connected to supply mains that include protective earth connection WARNING Avoid spilling liquids on the Battery Back Up box or next to the base Risk of electric sho...

Page 10: ...power cord connection on the pole should be in line with the set screw as shown Once the pole is properly positioned securely tighten the set screw using the 1 8 hex wrench provided 3 Plug cord from upright into Battery Box outlet 4 Do not install the Arm Assembly with the Light Head attached Installation with the Light Head attached can cause damage to the Light Refer to the product manual for at...

Page 11: ... Clear visual access to the indicator lights A distance between the Battery Backup unit and the light that is less than the supplied cable Ample room around the box for ventilation Wall stud or other secure mounting technique that accommodates a secure mounting of the box to the wall The Ceiling Wall Mounted Battery Back Up should be mounted based on the direction of a licensed structural engineer...

Page 12: ... 4 Step 1 To prevent a shock hazard turn off the mains input to the arm Disconnect and remove the light head from the arm Remove the arm and locate the power supply in the power supply enclosure on the arm Remove the Power Supply enclosure cover then remove the entire plate including the power supply by Cutting the Molex connector Cut the red and black DC wires that run through the arm Remove the ...

Page 13: ...the power supply housing in the orientation shown to the right Step 3 Connect the red and black DC wires that run though the arm to the terminal block as shown Be sure that the Red Wire connects to the Red Wire and that the Black Wire connects to the Black Wire Secure the plate to the housing by installing the four mounting screws The two ground wires should be attached using the mounting screws R...

Page 14: ... rod Connect the cable assembly to the DC output block inside the battery backup Connect mains to the mains input block inside the battery backup unit Refer to the light manual for the remainder of installation instructions Step 4 Wall Mount Feed the supplied cable assembly through the Wall Mount Bracket and into the Power Supply Housing Connect the cable assembly to the DC output block inside the...

Page 15: ...tic ceiling casting cover and locate the power supply plate Disconnect the mains connectors for the power supply you want to bypass These connectors can be found on the mains power distribution block Disconnect DC wires coming from the power supply at the DC output power distribution block on the ceiling casting A connection must be made between the DC power distribution block on the ceiling casti...

Page 16: ...the Battery Back Up you must first connect the 24VDC Battery to the system To do this remove the cover panel and then remove the RED insulating cover from the battery connector Once the insulator is removed connect the battery to the system by connecting the two gray connectors Once this is done connect the unit to mains and replace the cover The AC Power and Battery Charging LED s will light Remo...

Page 17: ...ill light when the AC power supply is powered The Blue CHARGING BATTERY LED will light whenever the battery charger is delivering charge to the battery The Amber BATTERY IN USE LED will light whenever the AC power is unavailable and the battery is available to power the light head The Yellow LOW BATTERY LED will light when the Battery falls below 22V At this point an alternative light should be us...

Page 18: ...sping the upright pole gently pull back to tip the entire unit Use the foot on the ground to prevent backwards motion of the entire unit 3 While maintaining manageability of the unit gently roll the unit over the threshold 4 Once the threshold has been cleared gently return the front caster wheels to their resting position on the ground 5 To guide the rear caster wheel over the threshold simple ro...

Page 19: ...zard Figure 4 Battery Back Up Enclosure 2 Using a small flat blade screwdriver open the hinged entry door by inserting the blade of the screwdriver into the slot located at the top portion of the module Gently pull the door backward to open it With the same screwdriver gently pry the fuse holder loose then slide it out to expose the two fuses These fuses protect the Battery Back up circuitry and f...

Page 20: ...s enclosure If the fuse s need to be replaced see the following table for DC Power 5 If the unit should continue to blow fuses do not use the unit contact the factory immediately CAUTION To Reduce the Risk of Fire Replace the fuse s only with the same type and electrical rating as indicated on the fuse replacement label Fusing Specifications MI 750 MI 1000 System Two LED A C Power 1 5A 2 5A 2 0A D...

Page 21: ...l Inc does not warranty any damage or harm as a result of improper installation or maintenance If the battery in your unit needs to be replaced contact your distributor and order Medical Illumination International Part No 0001824 24 Volt DC Rechargeable Sealed Lead Acid Battery Be prepared to provide the Model No Serial No and the exact problem experienced with the equipment To remove the battery ...

Page 22: ... to damage the wiring or other components when removing or replacing the battery 6 Install the new battery into the compartment and fasten the battery to the enclosure 7 Connect the battery connector to the connector in the housing 8 Connect the output cable to the power outlet receptacle and run the light on battery power only to make sure the unit is functioning properly 9 Fasten the metal cover...

Page 23: ... Wall unit Fuse is blown Plug in unit to AC source Reset circuit breaker Ceiling Wall version only Replace Fuse Light does not illuminate when power fails Failure of the Fuse or Battery Contact Medical Illumination Customer Service for service Battery Back Up power only lasts a short time Battery is dead defective Remove and replace the battery inside of the unit Battery does not charge or the AC ...

Page 24: ...y Assurance Manager REVISION HISTORY Rev Issue Date Summary of Changes Effectivity Date A 01 20 16 Initial Release 03 18 16 B 05 24 16 Add Battery Connector Removal in Section 4 05 25 16 C 02 20 19 Update CISPR statement on Agency Approval Page 02 20 19 REVISION APPROVAL Title Signature Date Operations Manager Engineering Manager Quality Assurance Manager Original Signature Copy on File ...