MCScap. User manual 

 

Medical Computer Systems Ltd.© 

tel. +7 495 9133194 

Passage 4922, 4-2, Zelenograd,  

fax. +7 495 9133195 

Moscow, 124460, Russian Federation 

www.mks.ru

 

Page 11 from 16 

Parameter 

Value 

SET 

CLINIC 

SLEEP 

PROFESSIONAL 

ELECTRODE SET 

Dielectric strength of 
electrodes insulation 

1500 V 

1500 V 

1500 V 

1500 V 

Diameter of a conductive 
surface 

6(+1/-1) mm 

3(+1/-1) mm 

6(+1/-1) mm 

depending on the type 

of electrode 

Square of a conductive 
surface 

26(+1/-1) mm

2

 

7(+1/-1) mm

2

 

≥ 30 mm

2

 

depending on the type 

of electrode 

The impedance of the 
electrode 

≤ 5 kΩ 

≤ 5 kΩ 

≤ 5 kΩ 

≤ 5 kΩ 

Spare electrode 

no

 

no

 

Channels in Standard caps 

Fp1, Fp2, F3, F4, C3, 

C4, P3, P4, O1, O2, 

F7, F8, Т3, T4, T5, 

T6, Cz, Fz, Pz, GND 

Fp1, Fp2, F3, F4, C3, 

C4, P3, P4, O1, O2, 

F7, F8, Т3, T4, T5, T6, 

Cz, Fz, Pz, GND 

Fp1, Fp2, F3, F4, C3, 

C4, P3, P4, O1, O2, 

F7, F8, Т3, T4, T5, T6, 

Cz, Fz, Pz, GND 

According to system 

10-20, 10-10, 10-5 

Ear electrodes 

no 

no, 

с

an be included by 

order 

no 

can be included by 

order

 

yes 

Marking of the electrodes 

yes 

yes 

yes 

no  

(labels for marking are 

Included) 

Additional wire protection 

no 

no 

yes 

no 

Exit of electrode cable from 
the cap 

back of the head 

top of the head 

back of the head 

no 

Connector type 

D-sub DB25 Male 

D-sub DB25 Male 

D-sub DB25 Male 

TouchProof 1.5 mm 

(DIN 42 802-ST) 

Summary of Contents for MCScap Clinic

Page 1: ...1993 2018 Medical Computer Systems Ltd MCScap EEG electrode caps USER MANUAL...

Page 2: ...fic researches GMDN code 63131 Electroencephalographic electrode cap reusable Product code by 93 42 EEC MD 0100 General non active non implantable medical devices Classification class I as per Appendi...

Page 3: ...s to any part of the product including cables specified by the manufacturer Otherwise degradation of the performance of this equipment could result If such use is necessary this equipment and the othe...

Page 4: ...d service Please look thoroughly through this manual before its use The manual contains detailed information and guidelines required for proper and safe operation of this product Failure to follow the...

Page 5: ...belt Reusable EEG electrodes have a channel for gel and the recording surface is silver silver chloride Ag AgCl Sintered technology of sensors guaranties minimal polarization and long term stability E...

Page 6: ...and Inion The frontopolar and occipital electrode positions should now be placed correctly Otherwise choose another cap size When measured correctly the frontopolar electrodes e g Fp1 Fp2 should lie d...

Page 7: ...ozzle dispenser should sit tightly on the syringe and should not hang or subside Use only supplied the nozzle dispenser with blunt needle Spiky needle can damage patient skin Slide the hair directly u...

Page 8: ...pull the electrode Under most circumstances cleaning of caps and electrodes is carried out in a detergent Be attentive dish detergents often leave a film A children s shampoo will be a good choice Im...

Page 9: ...nectors do not comply with the cap Replace the electrode or the cap Electrode connectors can not be connected to the electroencephalograph Electrode connectors do not comply with the electroencephalog...

Page 10: ...scientific research high resolution EEG MR compatible EEG available by order routine EEG scientific research Electrode type MCScap MCScap T MCScap NTC MCScap NT available by order MCScap E MCScap T M...

Page 11: ...ng on the type of electrode The impedance of the electrode 5 k 5 k 5 k 5 k Spare electrode no no 1 1 Channels in Standard caps Fp1 Fp2 F3 F4 C3 C4 P3 P4 O1 O2 F7 F8 3 T4 T5 T6 Cz Fz Pz GND Fp1 Fp2 F3...

Page 12: ...G starter kit gel syringe needle brush bag User manual 2 textile caps MCScap 10 20 with fixing rings 26 EEG electrodes MCScap E ear electrode fixators 2 sets of labels for marking electrodes EEG start...

Page 13: ...ghted in bold font 2 Cycle refers a use disassembly washing and disinfection of the helmet cleaning and disinfection of the electrodes Compliance Restriction of Hazardous Substances The object is in c...

Page 14: ...460 Russian Federation www mks ru Page 14 from 16 10 20 system Electrode positions and labels yellow circles indicate positions of the 10 20 system T3 T4 T5 and T6 corresponds to T7 T8 P7 and P8 of th...

Page 15: ...n Federation www mks ru Page 15 from 16 10 10 system Electrode positions and labels yellow circles indicate positions of the original 10 20 system white circles indicate additional positions introduce...

Page 16: ...MCScap size table Correct size is important for quality EEG acquisition and patient comfort Check the head circumference of patient before EEG examination...

Reviews: