Technical data
Parameter
Value
SET
CLINIC
SLEEP
PROFESSIONAL
ELECTRODE SET
Image
Recommended
application area
routine EEG
routine EEG,
sleep EEG,
high resolution EEG,
TMS-EEG
scientific research,
high resolution EEG,
MR-compatible EEG
available by order
routine EEG,
scientific research
Electrode type
MCScap-С
MCScap-T
MCScap-NTC
MCScap-NT available
by order
MCScap-E
MCScap-T, MCScap-
NT, MCScap-NTH
available by order
Material of electrode sensor
Ag/AgCl
Ag/AgCl sintered
Ag/AgCl sintered
Ag/AgCl sintered
Number of electrodes
(channels)
20 (19)
20 (19)
from 11 up to 129
available by order
20 (19)
from 11 up to 129
available by order
25 (24)
from 11 up to 129
available by order
Electrode polarization
≤ 50 mV
≤ 50 mV
≤ 50 mV
≤ 50 mV
Resistance of electrodes
insulation
≥ 1000 MΩ
≥ 1000 MΩ
≥ 1000 MΩ
≥ 1000 MΩ