5 Mounting
5.1 Mounting of the Table adapter with connecting arm
To mount the
Table adapter with connecting arm
1 Position the connecting arm assembly with the locking lever at the center of the
connecting bar/rod.
2 Place the adjustable end bracket onto the left end of the connecting tube.
3 Loosen the screws of the connecting arm
4 Connect the transitional member to the connecting arm assembly.
5 Fix the screws.
5.2 Mounting to operating table, adjustments
Mounting the Table adapter with connecting arm to the operating table:
1 Push the two end brackets into the appropriate receptacles of the operating table or
directly to OR rail
2 Proceed with step 4 if you do not need to adjust the Table adapter with connecting arm.
3 Adjust the left end bracket in such a way as to account for the distance between the
two holders at the operating table. Use the hex wrench which is located in the
connecting tube:
a) Loosen the adjustment screw at the left end bracket by means of the wrench
b) Turn the adjustment screw counterclockwise one full turn and carefully slide the end
bracket over the connecting tube to the desired position.
4 Turn the receptacle fastening screws at the operating table until fully tightened.
5 Position the adjustment screw of the left end bracket into the groove in the
connecting tube (otherwise, the Table adapter with connecting arm will not be rigid).
Carefully tighten the adjustment screw.
6 Move the connecting arm assembly to the desired position and fix the screws at
the connecting arm assembly.
Important:
The adjustable left end bracket allows you to easily mount the Table adapter with
connecting arm to standard operating tables.