Equipment Specifications
- 28 -
5.5 General
Display:
Two 3-digit, 7-segment LED
characters 1.0 cm. (0.4 in.) in height
Dimensions:
19.1 (L) x 8.9 (W) x 3.5 (H) cm.
7.5 (L) x 3.5 (W) x 1.4 (H) in
Weight:
516 g
18.2 oz
5.6 Sensors
Mediaid Opto-Plethysmographic pulse oximetry and temperature sensors with an
RJ12 connector and extension cable lengths 1.2 or 2.4 m / 4 ft or 8 ft Red Light
wavelength 660 nm @ 565
µ
W Infrared Light wavelength 910 nm @ 1.8 nW
5.7 Power Source
NiCad rechargeable battery (12 hours of operation) or power adaptor input
(100 – 250V AC, 50/60 Hz, 0.5A). Power Adaptor Output - 15V DC 1.0A
5.8 Environmental Conditions
Atmospheric Pressure:
1026 – 377 hPa
(770 – 282.45 mm Hg)
Relative Humidity:
5 to 95% (noncondensing)
Operating Temperature:
32 – 104° Fahrenheit (0 – 40° Celsius)
Storage/Transport
Temperature:
-22 – l49° Fahrenheit (-30 – 65° Celsius)
5.9 Equipment Classification
The Mediaid Model 340V Pulse Oximeter is classified according to CAN/CSA
C22.2 No. 601.1, EN 60601-1, and UL2601-1:
Type of protection against electric shock:
CLASS 1
and
INTERNALLY POWERED
.
TYPE BF APPLIED PARTS
.
Degree of protection against harmful ingress of water:
IPX 0 ACCORDING TO IEC 529
.
Degree of safety of application in the presence of a flammable
anesthetic mixture with air, or with oxygen or nitrous oxide:
THIS EQUIPMENT IS NOT SUITABLE FOR USE IN THE PRESENCE
OF A FLAMMABLE ANESTHETIC MIXTURE WITH AIR, OR WITH
OXYGEN OR NITROUS OXIDE.
Mode of operation:
CONTINUOUS OPERATION
Summary of Contents for 340V
Page 2: ......
Page 4: ...ii...
Page 10: ...Principe d Utilisation 4...
Page 11: ...Keys Indicators and Symbols 5 Chapter 2 Keys Indicators Symbols 2 1 Keys and Indicators...
Page 30: ...Principe d Utilisation 24...
Page 32: ...Principe d Utilisation 26...
Page 38: ...Principe d Utilisation 32...
Page 40: ......