
Operating the Model 340V
- 17 -
Silence alarms with one of the following actions:
•
Silence the alarms for a period of 60 seconds with one short
depression of the ALARM OFF KEY. The ALARM OFF INDICATOR
illuminates and remains on constantly until the alarms are
reactivated again.
•
Completely disable the audible alarm with one long
depression of the
ALARM OFF KEY. The LED displays show
A L O FF
, the ALARM OFF
INDICATOR begins to flash, and the Model 340V reverts to normal
monitoring.
Reactivate alarms with one of the following actions:
•
Reactivate silenced alarms with one short
depression of the ALARM
OFF KEY. The ALARM OFF INDICATOR is no longer lit.
•
Reactivate disabled alarms with one long
depression of the ALARM
OFF KEY. The ALARM OFF INDICATOR stops flashing. The LED
displays show
A L ON
.
3.7 Power
The Model 340V can be powered by
either a power adaptor or the internal,
rechargeable NiCad battery. Battery
operation alone will provide the Model
5340V with approximately 12 hours of
operation. Once electric power is
applied, the battery charges regardless
of the operational state of the Model
5340V.
WARNING
All Mediaid Inc. pulse
oximeters are designed to be
used with the power adaptor
provided by Mediaid Inc. at the
time of purchase. The use of
any other adaptor may damage
the pulse oximeter and will
VOID the Mediaid Inc.
Warranty. Refer to the product
number of the power adaptor
when reordering a new
adaptor. Include the serial
number of the Model 340V as
this allows Mediaid Inc. to
determine the adaptor
required.
WARNING
Avoid excess tension on the
adaptor cable for safe and
continuous operation.
Summary of Contents for 340V
Page 2: ......
Page 4: ...ii...
Page 10: ...Principe d Utilisation 4...
Page 11: ...Keys Indicators and Symbols 5 Chapter 2 Keys Indicators Symbols 2 1 Keys and Indicators...
Page 30: ...Principe d Utilisation 24...
Page 32: ...Principe d Utilisation 26...
Page 38: ...Principe d Utilisation 32...
Page 40: ......