Troubleshooting
13 Troubleshooting
Motor not running
Check if:
– the Clario is switched on�
– the mains voltage is correct and if the mains plug is inserted correctly into the fixed mains
socket�
– the internal battery is charged (only AC / DC version under battery operation)�
Running time from battery insufficient
Calibrate the rechargeable battery� See chapter 12�
Pump is too loud
Check the position and seal of the valve discs inside the SafetyChamber�
Insufficient vacuum
Check if:
– the vacuum regulator is set correctly�
– the tubings are not defectiv e or broken� If necessary, replace�
– all plug-in connections are tight�
– the suction canister / lid have no cracks, brittle areas, discolouration� If necessary, replace�
– the seal of the valve discs inside the SafetyChamber is tight�
– the membrane is clean and has no tears�
No LED lit
(AC / DC version only)
The rechargeable battery is fully charged�
LED lit up
(AC / DC version only)
– The rechargeable battery is being charged�
– Pump is plugged in�
LED blinks slowly
(AC / DC version only)
The rechargeable battery is soon empty� The battery power is sufficient for about 10 minutes
of pumping�
LED blinks rapidly
(AC / DC version only)
The rechargeable battery is defective� For replacing the defective battery see chapter 10�