Battery calibration
CAUTIONS
Calibrate the rechargeable battery before first use
and after storage duration > 2 months� Replace
battery if the calibration is not successful�
Calibrating NiMH rechargeable battery
Pull out the mains plug from the fixed mains socket�
2�
Connect the Clario Toni to a fixed mains socket�
The rechargeable battery is being charged�
5�
Wait until the charging indicator goes out�
6�
Repeat step 1 to 6� The Clario Toni AC / DC is then
ready to use�
7�
Let the pump run until the rechargeable battery is
empty and the pump comes to a stop�
4�
3�
Switch on the Clario Toni�
The pump will start running�
Make sure that the rechargeable battery is fully charged�
1�
Storage of the rechargeable battery
In order to keep the self-discharge at a minimum, store
the Clario Toni and the removed replacement recharge-
able batteries at temperatures below 25 °C (77 °F)�
Repeat the calibration process every 60–90 days�
Summary of Contents for Clario toni
Page 35: ...35 EN Disposal Ask your healthcare provider how to dispose of used tubings...
Page 143: ...143 IT Smaltimento Consultare l assistente sanitario per lo smaltimento di tubi...
Page 287: ...287 NO Avfallsh ndtering Kontakt helseinstitusjonen for r d om kasting av brukte slanger...
Page 359: ...359 PT Elimina o Pergunte ao prestador de cuidados de sa de como eliminar as tubagens usados...
Page 362: ...362 Notes...
Page 363: ...363 EN Notes...
Page 364: ...364 Notes...
Page 365: ......