61
Tab 2
Guidance and manufacturer’s declaration - electromagnetic Immunity
Immunity test
IEC 60601-1-2 Test level
Compliance level
Electrostatic discharge (ESD)
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast transient/burst
Power supply lines: ±2 kV
100 kHz repetition frequency
Power supply lines: ±2 kV
Surge
IEC 61000-4-5
line(s) to line(s): ±1 kV.
line(s) to line(s): ±1 kV.
100 kHz repetition frequency
Voltage dips, short interruptions and
voltage variations on power supply
input lines
IEC 61000-4-11
0% 0.5 cycle
At 0º, 45 º, 90 º, 135 º, 180 º, 225 º, 270 º
and 315 º
0% 1 cycle
And
70% 25/30 cycles
Single phase: at 0
0% 250 cycle
0% 0.5 cycle
At 0º, 45 º, 90 º, 135 º, 180 º, 225 º, 270 º
and 315 º
0% 1 cycle
And
70% 25/30 cycles
Single phase: at 0
0% 250 cycle
Power frequency magnetic field
IEC 61000-4-8
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz
Conduced RF
IEC61000-4-6
150KHz to 80MHz: 3Vrms
6Vrms (in ISM and amateur radio bands)
80% Am at 1kHz
150KHz to 80MHz: 3Vrms
6Vrms (in ISM and amateur radio bands)
80% Am at 1kHz
Radiated RF
IEC61000-4-3
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
10 V/m
80 MHz – 2,7 GHz
80 % AM at 1 kHz
NOTE UT is the a.c. mians voltage prior to application of the test level.
Guidance and manufacturer’s declaration - electromagnetic emissions
Emissions test
Compliance
RF emissions
CISPR 11
Group 1
RF emissions
CISPR 11
Class B
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/ flicker emissions
IEC 61000-3-3
Compliance
Tab 1