2. debritom
+
set-up
debritom
+
QuickCard - EN
This QuickCard does not replace the Instruction for use REF 9000.5503
Attach the
luer connection
- make sure
the connection is tighten securely
Attach
rinsing fluid
, connect the connecting tube to pump
Connecting tube
: close clamp/spike into rinsing fluid/open venting cover
Medaxis AG
Bahnhofstrasse 9
6340 Baar, Switzerland
www.medaxis.ch
© Medaxis
REF 9000.0000_d EN / 08.11.2017
1. Turn on debritom
+
Activate foot pedal
1x
to connect with
debritom
+
debritom
+
ON
2a
2b
2c
2d
3. Test of water jet
Press foot on/off switch
§
For a vusual test of water jet (flat or
point, water jet thickness)
§
To test the water jet intensity on your
hand – start with a distance of 25 cm
(approx. twice the hanpiece-length).
Guide the jet starting at the wound
edge, and move the jet slightly back
and forth.
§
If something doesn‘t work, replace the
handpiece
Intensity
Intensity
5. Intensity-Level
6. Patient change / end of the treatment
debritom
+
OFF
Remove connecting tube/ Remove pump and handpiece as one unit and dispose of it
6b End of the treatment
Close clamp
Remove handpiece and dispose of it
Attach LuerCap
6a Patient change, pump one day
For next patients
=> 2d
4. Optimal working parameters
Open
clamp
2e
TIP flat jet:
The flat jet should be applied like a spatula
to the wound.
debritom
+
handpiece standard jet
REF 2000.0003
Working distance recommended
8-15cm
debritom
+
handpiece flat jet narrow
REF 2000.0004
Working distance recommended
4-10cm
debritom
+
handpiece flat jet wide.
REF 2000.0005
Working distance recommended
3-8cm
Working angle:
3 – 45 degrees to wound surface
„click“
Insert pump
as shown until it stops / turn clockwise until click
Intensity levels at START for each handpiece : Level 4 (Level 1-5/weak-strong)