7. Error messages
NOTE:
The on/standby button lights up red as an error message occurs. This button can be used to acknowledge the fault after rectification.
Battery foot on/off switch almost empty
Troubleshooting: Replace batteries (Battery type 2 x LR03 / AAA, chapter 11)
Pump in use for too long time period
Troubleshooting: Replace pump
Handpiece in use for too long or trying to use with different single use pump
Troubleshooting: Replace handpiece
Handpiece missing or handpiece not recognized
Troubleshooting: Replace handpiece. If the error can not be resolved, contact the Medaxis service department immediately.
Nozzle blocked or motor overheated
Troubleshooting: Change handpiece and pump if necessary. If the error can not be resolved, contact the Medaxis service department
immediately.
Pump not inserted correctly
Troubleshooting: Replace pump and reinsert it correctly, or replace it.
General system error
Troubleshooting: Please contact the Medaxis service department immediately.
No foot on/off switch detected
Troubleshooting: Press the foot on/off switch. If foot on/off switch cannot be detected, re-pair. (See chapter 11)
press
confirm
Search for Bluetooth device for pairing
Pressing the on/standby and ‚+‘ buttons for > 1 s at d will show the ‚Search for Bluetooth device‘ symbol for pairing a new or
replacement of the foot on/off switch.
Bluetooth signal of foot on/off switch detected - confirm
As soon as debritom
+
has detected the Bluetooth signal of the foot on/off switch, this symbol appears on the display.
debritom
+
connected with foot on/off switch
This symbol appears on the display as the foot on/off switch is connected to debritom
+
.
Intensity levels
The intensity levels can be adjusted between 1 - 5. (1 weakest/5 strongest)
8. Symbols on debritom
+