9
Aspiration
Comply with the relevant hygiene regulations when sep-
tic fluids are aspirated. Only use the aspirator in a vertical
position.
• Check aspirator for visible damage prior to use, e.g.
ruptures or cracks in tubes or septic fluid containers.
• Replace defective parts.
• Position the aspirator in a safe position.
• Use the rocker switch (3) to switch on the aspirator.
• Attach finger tip or suction element.
If the aspirator is used without regulating unit, it is man-
datory that a finger tip be inserted in the suction tube.
• Adjust vacuum: Close the finger tip control (4) with a
finger, for continuous operation close suction opening
with the plug.
• Aspirate septic fluid.
If frothing occurs during aspiration, use standard froth
preventers since otherwise the overflow prevention de-
vice (12) might not close in the case of excess aspiration.
When the septic fluid jar is filled to two thirds
(maximum) switch off the device and empty the septic
fluid jar.
If septic fluid enters the aspirator, switch it off immedia-
tely and have it cleaned by the Technical Service; if
required, the Technical Service will also replace compon-
ents.
• If the overflow prevention device (12) was closed, the
suction unit needs to be replaced.
• Re-prepare aspirator after use.
!
)