Aspiration

Comply with the relevant hygiene regulations when sep-
tic fluids are aspirated. Only use the aspirator in a vertical
position.

• Check aspirator for visible damage prior to use, e.g.

ruptures or cracks in tubes or septic fluid containers.

• Replace defective parts.

• Position the aspirator in a safe position.

• Use the rocker switch (3) to switch on the aspirator.

• Attach finger tip or suction element.

If the aspirator is used without regulating unit, it is man-
datory that a finger tip be inserted in the suction tube.

• Adjust vacuum:  Close the finger tip control (4) with a

finger, for continuous operation close suction opening
with the plug.

• Aspirate septic fluid.

If frothing occurs during aspiration, use standard froth
preventers since otherwise the overflow prevention de-
vice (12) might not close in the case of excess aspiration.

When the septic fluid jar is filled to two thirds
(maximum) switch off the device and empty the septic
fluid jar.

If septic fluid enters the aspirator, switch it off immedia-
tely and have it cleaned by the Technical Service; if
required, the Technical Service will also replace compon-
ents.

• If the overflow prevention device (12) was closed, the

suction unit needs to be replaced.

• Re-prepare aspirator after use.

!

)

Summary of Contents for 57505128

Page 1: ...1 Septic Fluid Aspirator P 7050 P 7050 100 01 57505128 REF 57520525 Valid as of August 03 Operating Instructions Sekretsauger P 7050...

Page 2: ...sories 7 Operation 8 Starting the operation 8 Preparation of accessories 8 Aspiration 9 Cleaning disinfection sterilization 10 Troubleshooting 12 Maintenance 12 Accessories Consumables Spare Parts 13...

Page 3: ...atients and operating staff as well as for the prevention of damage to the device and or functional defects The meaning of the pictorial signs is as follows STOP Danger for patient and operating staff...

Page 4: ...ple 7 O Ring 8 Gasket 9 Septic fluid jar 10 Suction unit 1 3 2 4 5 6 7 8 10 9 Humidity filter Sekretsauger P 7050 Accessories Manometer with overflow prevention device 11 Manometer 12 Overflow prevent...

Page 5: ...tion according to instruction by qualified personnel If other than original parts are used performance and safety might be impaired The device must be used in vertical position only Do not switch on t...

Page 6: ...e P 7050 aspirator the vacuum is generated by an electrically driven pumping aggregate The vacuum is 0 8 bar and can be regulated with a ma nometer with regulating valve and overflow prevention device...

Page 7: ...contai ner the tube nipple for the connection to the vacuum source must fit for the connecting tube Suction tube Vacuum resistant to 0 8 bar must not collapse inner diameter of tube must correspond t...

Page 8: ...via the hu midity filter to the aspirator black tube nipple Pay attention to direction of flow The side with the text must point in the direction of the suction unit Plug in power plug Carry out a fu...

Page 9: ...be Adjust vacuum Close the finger tip control 4 with a finger for continuous operation close suction opening with the plug Aspirate septic fluid If frothing occurs during aspiration use standard froth...

Page 10: ...ot treat septic fluid containers made of polysulphone with strong acids or bases Risk of fractures due to tension Always disassembly the overflow prevention device 12 for preparation Unscrew it from t...

Page 11: ...e are only valid for original Heraeus Med components If you use components of other manufacturers follow the procedures of the respec tive manufacturer Make sure that no fluid penetrates the aspirator...

Page 12: ...eck fuses Power plug does not correctly fit Insert power plug correctly Power cable defective Motor does not run Septic fluid jar cap not correctly placed Place septic fluid jar cap correctly Humidity...

Page 13: ...630 Spare parts This list contains spare parts which can be replaced by the final user A com plete spare part list will be supplied to especially trained and qualified personnel Septic fluid jar 0 7 l...

Page 14: ...on Transportation storage conditions Temperature 15 C to 50 C Rel air humidity 100 rel humidity Operating conditions Temperature 15 C to 30 C Air humidity 30 to75 rel humidity Performance data Art no...

Page 15: ...cture of P 7050 materials are used which are utilizable or can be recycled and which do not carry a special or extraordinary risk during disposal Heraeus Med will recycle a device which has become un...

Page 16: ...Rastatt D 76437 Rastatt AUSGABE AUGUST 2003 SECTION AUGUST 2003 REF 5752 0525 REF 5752 0525 VERTRIEB 0 18 03 21 21 33 SALES 49 72 22 932 644 VERTRIEB FAX 0 18 03 21 21 77 FAX 49 72 22 932 571 KUNDEND...

Reviews: