Troubleshooting
Troubleshooting / Maintenance
Maintenance
We recommend to have the device inspected by trained
and qualified personnel at least once a year.
Repair work can be done by personnel trained by Heraeus
Med, or call the Heraeus Med repair service. The manufa-
crurer will supply spare part lists and other documents
to trained and qualified personnel for parts which have
been assessed as being repairable.
Before returning a device for repair, clean and disinfect it
first.
!
Defect
Reason
Corrective action
No power
Check fuses
Power plug does not correctly fit
Insert power plug correctly
Power cable defective
Motor does not run
Septic fluid jar cap not correctly placed
Place septic fluid jar cap correctly
Humidity filter clogged
Exchange filter
Rupture in tube
Exchange tube
Seal dirty
Exchange seal
Seal porous
Exchange seal
Suction element is clogged
Clean or exchange suction element
Have device repaired
Hardly any / no displacement
Aspirator does not run
Overflow prevention device contaminated
with septic fluid
Remove overflow prevention device and clean
it
Septic fluid container full,
overflow prevention device closed
Tube connection in septic fluid jar cap
is clogged
Empty suction unit
and clean or replace it
Clean connection