Medacta Shoulder System Manual Download Page 7

7

6.  BONE GRAFT SHAPING

The bone graft should be shaped according to the planned/

desired dimensions.

Select the Graft Shaper according to the desired angulation 

and assemble it with the Graft Holder. Four different angles 

are available: 0°, 10°, 20° and 30°.

10.

Put the bone graft into the assembled instrument making 

sure to place the curved surface of the graft on the bottom 

of the device.

Rotate the knob to adjust the graft height according to the 

desired dimensions. The numbers on the Graft Shaper 

indicate the bone graft thickness. Depending on the chosen 

angle, the numbers indicate the thickness of the two sides 

of the graft.

Then, press the lateral pincer in order to firmly hold the graft 

in place. Use a HEX 3.5 screwdriver to tighten the screw 

inside the pincer to lock it.

11.

Perform the final shaping using an oscillating saw

12.

CAUTION

 

 

Pay attention to keep the hands away from the cutting 

plane while performing the resection.

Part numbers subject to change.

NOTE FOR STERILIZATION

The instrumentation is not sterile upon delivery. Instruments must be cleaned before use and sterilized in an autoclave respecting 

the US regulations, directives where applicable, and following the manufactures instructions for use of the autoclave.  

For detailed instructions please refer to the document “Recommendations for cleaning decontamination and sterilisation of 

Medacta International orthopaedic devices” available at www.medacta.com.