4.  BONE GRAFT REAMING

Slide the Central Piston Guide over the K-wire and place it 

on the reamed humeral head surface. Use a HEX 3.5 

screwdriver to tighten the screw inside the instrument in 

order to lock it on the K-wire.

3.

The Central Piston Guide is available in two different height 

options, depending on the desired reaming depth.

Connect the Modular Trephine - Graft Reamer with the 

Modular Trephine - K-wire Adaptor.

4.

Assemble the Reamer Handle with the T-handle and 

connect them to the previously assembled instruments.

5.

Slide this construct on the K-wire and the Central Piston 

Guide. Ream the humeral head until the desired depth is 

reached. The numbers on the external surface of the Graft 

Reamer represent the height of the bone graft.

If the Central Piston Guide is used as a mechanical stop, 

the maximum bone reaming depth is 10mm or 20mm, 

depending on the selected Central Piston Guide. If a higher 

bone graft height is necessary, unscrew the Central Piston 

Guide, remove it and proceed with the reaming.

6.

Extract the instruments.

Select the desired length of the Central Peg Reamer and 

connect it to the Reamer Handle. Slide the assembled 

reamer over the K-wire and use a power tool to ream until 

the mechanical stop is reached.

7.

Remove the instruments.

Alternatively, it is possible to perform the steps described 

above in only one phase.

In this case, connect the Modular Trephine – Central peg 

reamer to the Modular Trephine – Graft reamer and 

assemble them with the Reamer Handle and T-handle.