Maintenance
Cleaning
Frame Cleaning
- The frame can be cleaned/disinfected with a disinfecting wipe. Disinfectant
should always be applied to a cloth; never spray directly on frame. Always follow
manufacturer’s instructions for proper use of cleaning agent. After applying cleaning agent,
allow the
surface to dry prior to placing unit back in service. Clean frame as needed.
Pad/Cushion Cleaning
-
All cleaning solutions must be properly diluted according to the
manufacturer’s instructions. Please follow standard institutional wipe down and infection control
procedures.
Maintenance and Service
NOTICE: RECOMMENDED TO PERFORM PREVENTIVE MAINTENANCE EVERY 12
MONTHS FOR BEST PERFORMANCE.
1. Inspect the frame for weld cracks or tears.
2. Check that all bolts, pins and fasteners are still in working order.
3. Perform functional test on the chair.
4. Inspect the casters for wear and tear.
Preventative Maintenance Inspections
Problem Resolution
Inspect chair frame for cracks, tears or bending,
particularly at welds and joints.
Remove chair from service. Contact Med-Mizer®
customer service.
Check that screws, bolts, pins and other fasteners are
in place, secured and properly tightened.
Remove chair from service. Tighten loose bolts and
screws. Contact Med-Mizer® customer service to
replace missing or damaged fasteners.
Inspect gas springs for excessive wear and
misalignment.
Remove chair from service. Contact Med-Mizer®
customer service.
Inspect all gas spring cabling for wear and tear.
Remove chair from service. Contact Med-Mizer®
customer service to replace power cord.
Third adjustment
forward
Backrest and
cylinder location
Backrest and
cylinder location