www.tensmachineuk.com
38
Guidance and manufacturer
s declaration - electromagnetic emissions
The device is intended for use in the electromagnetic environment
specified below. The customer or the user of the device should
assure that it is used in such an environment.
Emissions test
Compliance
Electromagnetic environment -
guidance
RF emissions
Group 1
The device must emit electromag-
netic energy in order to perform its
CISPR 11
intended function. Nearby electronic
equipment may be affected.
RF emissions
Class B
The device is suitable for use in all
CISPR 11
establishments other than domestic
Harmonic emissions Class C
those directly connected to the public
IEC 61000-3-2
low-voltage power supply network that
Voltage fluctuations
Complies
supplies buildings used for domestic
/ flicker emissions
purposes.
IEC 61000-3-3
¤”› ד⁄—Œ”flÃ×—“
The device complies with current EMC regulations.
The radio frequency emissions of the device are extremely low and in
all probability do not cause any interference with other devices in the
proximity.It is recommended that you do not place the device on top of
or close to other electronic devices.
Guidance and manufacturer
s declaration - electromagnetic immunity
The device is intended for use in the electromagnetic environment specified
below. The customer or the user of the device should assure that it is used
in such an environment.
IMMUNITY test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply
input lines IEC 61000-
4-11
Power frequency
(50/60 Hz) magnetic
field IEC 61000-4-8
IEC 60601 test
level
!
6 kV contact
!
8 kV air
!
2 kV for power
supply lines
!
1 kV line(s) to
line(s) and neutral
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
3 A/m
Compliance level
!
6 kV contact
!
8 kV air
!
2 kV for power
supply lines
!
1 kV line(s) to
line(s) and neutral
<5 % U
T
(>95 % dip in U
T
)
for 0,5 cycle
40 % U
T
(60 % dip in U
T
)
for 5 cycles
70 % U
T
(30 % dip in U
T
)
for 25 cycles
<5 % U
T
(>95 % dip in U
T
)
for 5 s
Not applicable
Electromagnetic environment -
guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered
with synthetic material, the relative
humidity should be at least 30 % .
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of the
device requires continued
operation during power mains
interruptions, it is recommended that
the device be powered from an
uninterruptible power supply or a
battery.
Not applicable
NOTE U
T
is the a.c. mains voltage prior to application of the test level.