Warning
Problem
Solution
0
1
2
3 s
5 min
The DL-Coil and implant are
disconnected.
Position the DL-Coil over the implant.
SONNET 2 is in microphone
monitoring mode.
Switch SONNET 2 off and on again.
The DL-Coil is positioned
over the wrong implant.
Position the DL-Coil over
the correct implant.
The DL-Coil cable is defective.
If the DL-Coil cable fails, the
DL-Coil will automatically
power off after 5 minutes.
1. Replace the DL-Coil cable.
2. Reposition the DL-Coil over the implant.
3. Switch your SONNET 2 off.
4. Switch your SONNET 2 back on.
0
10 s
The DL-Coil has powered off.
1. Reposition the DL-Coil over the implant.
2. Switch your SONNET 2 off.
3. Switch your SONNET 2 back on.
Random signals
The DL-Coil cable is defective.
Replace the DL-Coil cable.
31