INSTALLATION AND MODE OF OPERATION
It is vital to ensure straight inlet and outlet sections. If this cannot be achieved, flow
straighteners must be used or alternatively the measuring cross-section must be reduced.
Fig. 8 Installation between elbows, valves and pumps (D = DN)
If several sensors are connected in series, the distance between the individual sensors
must be at least one sensor length. If two or more sensors are mounted side by side, the
minimum distance is 1 m.
Avoid any buildup of gas at the highest point in the pipeline. (Fig. 9)
Fig. 9 Installation at the highest point