background image

mag-flux T4 

Magnetic flow meter

 

compact and separated

OPERATING INSTRUCTIONS 

© MECON GmbH BA / mag-flux T4 / EN / 19-04

Summary of Contents for mag-flux T4

Page 1: ...mag flux T4 Magnetic flow meter compact and separated OPERATING INSTRUCTIONS MECON GmbH BA mag flux T4 EN 19 04...

Page 2: ...MPRINT All rights reserved Any reproduction or usage without written authorisation of MECON GmbH even partial content is strictly prohibited Subject to change without notice Copyright 2019 by MECON Gm...

Page 3: ...DATA 20 5 1 Reference conditions 20 5 2 Flow sensor transducer 20 5 3 Condition of fluid 21 5 4 Transducer M1 21 5 5 Dimensions coupling connection compact design 22 5 6 Dimensions coupling connectio...

Page 4: ...s solely responsible for the suit ability intended use and corrosion resistance of the selected materials In particular it must be ensured that the materials selected for the wetted parts of the flow...

Page 5: ...6 1 3 Manufacturer s safety instructions The manufacturer is not liable for damages of any kind caused by the use of the equip ment including but not limited to direct indirect incidental punitive and...

Page 6: ...therwise agreed To prevent injury to the user or damage to the unit it is necessary that you read these operating instructions carefully before starting to use the equipment The operating instructions...

Page 7: ...7 MECON GmbH BA mag flux T4 EN 19 04 DEVICE DESCRIPTION 2 DEVICE DESCRIPTION 2 1 Scope of delivery mag flux T4 flow meter Operating Instruction Calibration certificate Fig 1 Scope of delivery...

Page 8: ...x M1 transducer 2 2 Nameplate Nameplate mag flux T4 sensor Fig 3 Nameplate mag flux T4 sensor Equipment type Order code Description code Options Options Serial no Serial number Connection Process conn...

Page 9: ...r product E v x B In a tubular measuring tube provided with an electrically insulating lining and where a liquid flows through with a velocity v and flow Q a magnetic field B is generated per pendicul...

Page 10: ...may only be used in combination with VdS approved pipe couplings manufactured by Anvil all Gruvlok mechanical grooved couplings Jinan Meide cast ing couplings type 1G Minimax Modgal Tyco Grinnell Mec...

Page 11: ...rtical pipelines Where there is a free pipe outlet the sensor should not be installed in a section of pipe that can run empty e g downpipes Fig 6 For installations in a downpipe ensure that the pipeli...

Page 12: ...the measuring cross section must be reduced Fig 8 Installation between elbows valves and pumps D DN If several sensors are connected in series the distance between the individual sensors must be at le...

Page 13: ...ional earth must be observed Potentials The signal outputs process outputs and the mains supply of the mag flux M1 transducer are galvanically insulated from each other and from the measuring circuit...

Page 14: ...wise no liability whatsoever is accepted for any resulting consequential damages If there is a fault the service department at MECON GmbH is available to help you Tel 49 0 2237 6 00 06 0 Fax 49 0 2237...

Page 15: ...sed in signals that are defined here These terms are used in the sense of the documentation and the references to the products themselves and have the following meaning Danger This means that death se...

Page 16: ...nalogue output Terminal Designation Function 1 PN Protective conductor 2 N Mains 3 L Mains phase 4 Impulse Pulse output passive 5 Impulse Pulse output passive 6 Status Status output passive 7 Status S...

Page 17: ...n Fig 12 Sensor connection of the mag flux M1 sensor Pin wire colour Connection cable Function 1 brown 24 V 2 white not used 3 blue Earth 0 V 4 black not used 5 grey Analogue output 4 20 mA Pin wire c...

Page 18: ...ducer housing is mounted in such a way that the display can be read in the correct position with a horizontal installation position and a direction of flow from left to right To change the orientation...

Page 19: ...nical means photocopying or otherwise duplicated or repro ducted by any other means without the prior written consent of MECON GmbH Great care has been taken in the development and preparation of this...

Page 20: ...0 12 Flange connection EN 1092 1 DN 15 DN 300 Measuring accuracy Measurement error Repeatability 0 5 of the measured value from 1 m s to 10 m s 0 4 of the measured value 1 mm s of 1 m s 0 15 of the me...

Page 21: ...tainless steel Hastelloy Corrosion protection class C2 slightly polluted atmosphere dry climate Housing material Die cast aluminium housing Electrical data Auxiliary energy Power input 230 V AC 15 10...

Page 22: ...m mm W mm D mm H mm 50 2 60 3 150 0 80 140 57 65 2 76 1 150 0 80 155 63 80 3 88 9 150 0 80 170 70 100 4 114 3 200 0 120 210 86 125 5 139 7 200 0 120 240 98 150 6 168 3 200 0 120 285 117 200 8 219 1 30...

Page 23: ...50 2 60 3 150 0 2 0 80 140 57 65 2 76 1 150 0 2 0 80 155 63 80 3 88 9 150 0 2 0 80 170 70 100 4 114 3 200 0 2 0 120 210 86 125 5 139 7 200 0 2 0 120 240 98 150 6 168 3 200 0 2 0 120 285 117 200 8 219...

Page 24: ...0 80 130 53 25 1 200 0 2 0 80 130 53 32 1 200 0 2 0 80 130 53 40 1 200 0 2 0 80 130 53 50 2 200 0 2 0 80 140 57 65 2 200 0 2 0 80 155 63 80 3 200 0 2 0 80 170 70 100 4 250 0 2 0 120 210 86 125 5 250 0...

Page 25: ...80 130 53 25 1 200 0 2 0 80 130 53 32 1 200 0 2 0 80 130 53 40 1 200 0 2 0 80 130 53 50 2 200 0 2 0 80 140 57 65 2 200 0 2 0 80 155 63 80 3 200 0 2 0 80 170 70 100 4 250 0 2 0 120 210 86 125 5 250 0 2...

Page 26: ...mm W mm D mm H mm G 14 0 150 0 2 0 80 130 53 G 19 0 150 0 2 0 80 130 53 G 1 27 0 150 0 2 0 80 130 53 G 1 33 0 150 0 2 0 80 130 53 G 1 38 0 150 0 2 0 80 130 53 G 2 48 5 150 0 2 0 80 140 57 Tab 10 Dime...

Page 27: ...f the sensor housing l L mm mm W mm D mm H mm G 14 0 150 0 2 0 80 130 53 G 19 0 150 0 2 0 80 130 53 G 1 27 0 150 0 2 0 80 130 53 G 1 33 0 150 0 2 0 80 130 53 G 1 38 0 150 0 2 0 80 130 53 G 2 48 5 150...

Page 28: ...H BA mag flux T4 EN 19 04 DESCRIPTION CODE 5 11 Dimensions of transducer mag flux M1 Fig 21 Dimensions of transducer mag flux M1 Weight 3 9 kg 8 6 lb 6 DESCRIPTION CODE The description code is made up...

Page 29: ...1092 1 DN 50 PN 40 0 1 178 l min 0 310 USgpm H D065 Flange connection EN 1092 1 DN 65 PN 16 0 1 990 l min 0 525 USgpm H H065 Flange connection EN 1092 1 DN 65 PN 40 0 1 990 l min 0 525 USgpm H D080 F...

Page 30: ...e connection ANSI B16 5 2 150RF 0 1 178 l min 0 310 USgpm H A065 Flange connection ANSI B16 5 150RF 0 1 990 l min 0 525 USgpm H A080 Flange connection ANSI B16 5 3 150RF 0 3 015 l min 0 800 USgpm H A1...

Page 31: ...DN 100 PN 40 0 4 712 l min 0 1 250 USgpm P D125 Flange connection EN 1092 1 DN 125 PN 16 0 7 363 l min 0 1 950 USgpm P H125 Flange connection EN 1092 1 DN 125 PN 40 0 7 363 l min 0 1 950 USgpm P D150...

Page 32: ...e connection ANSI B16 5 2 150RF 0 1 178 l min 0 310 USgpm P A065 Flange connection ANSI B16 5 150RF 0 1 990 l min 0 525 USgpm P A080 Flange connection ANSI B16 5 3 150RF 0 3 015 l min 0 800 USgpm P A1...

Page 33: ...separated with 5 m cable between sensor and transducer 1 230 V AC 50 60 Hz 15 VA 2 115 V AC 50 60 Hz 15 VA 3 24 V DC 15 15 W 1 without with cable gland M20 x 1 5 4 230 V 1 x circular connector M12 x...

Page 34: ...n accordance with these operating instructions Should you nevertheless need to return a device to MECON GmbH please observe the following points Attention For reasons of environmental protection and s...

Page 35: ...ication Tube empty Empty pipe detection is switched on Pipe is empty Ensure it is filled Coil current Line interruption to the field coil The signal outputs are set to zero Send mag flux T4 back to ME...

Page 36: ...36 MECON GmbH BA mag flux T4 EN 18 12 MECON GmbH R ntgenstr 105 50169 Kerpen Germany Tel 49 0 2237 600 06 0 Fax 49 0 2237 600 06 40 E Mail info mecon de Web www mecon de...