4. OPERATING INSTRUCTION

4.1 AC + DC Voltage Measurement

WARNING

Maximum input is 600V. Do not attempt to take any

voltage measurement that exceeds these limits.

Exceeding these limits could cause electrical shock

and damage to the clamp meter.

1. Set the rotary switch to the “ V ” position.

2. Insert the test leads in to the input jack. (Black to COM

and Red to V)

3. Connect the test leads in PARALLEL to the measured

circuit.

4. The  meter  will  automatically  switch  to  ACV  or  DCV

display.

5. The meter will automatically select the appropriate range.

6. Read  the  voltage  and  frequency  values  displayed  on

the LCD.

NOTE

The sensitivity for voltage frequency measurement is 1.2V

and the frequency range is 40 - 1KHz. If the frequency is

less than 40Hz the LCD may show Hz.

10. “   

 V 

 m

A ”Jack :

Connect  red  test  lead  for  Capacitance,  Resistance,
Diode,  Countinuity and  Micro  ampere  measurement
as a positive terminal.

11. 3 Phase Adapter :

Insert  plug  in  type  3  phase  adapter  in  terminals
“ COM ” and “ V ” for 3 phase 3 wire power measurement.

3. PARTS & CONTROLS

3.1 Description of Parts & Controls

2

5

4

10

8

9

3

7

6

1

11

Fig.

1

1.   Transformer jaws (+mark to face source) :

To pick up the AC current flowing through the conductor.

2.   Jaw opening trigger
3.  Data Hold (MX/MN) button :

a. Data Hold Mode :

Press it once to hold the measured value and store the 
value  in memory. Press again to release hold function. 
(except capacitance function)

b. Maximum and Minimum with Elapsed Time mode

0

1.  Select  desired  ACA,  ACV,  DCV,  C/ F  or  mA

function  and  perform  the  measurement,  until  the
measured value shows stable reading on the LCD.

2. Press  “  MX/MN  ”  button  for  2  seconds  to  enter

recording with Elapsed Time mode, LCD will show 
“  ®  ”  mark  to  lock  the  measurement  range  and 
“ MX/MN ” mark to indicate the current measured 
value  with  current  elapsed  time.  The  Auto  Power 
Off function will be automatically cancelled.

3. Press “ MX/MN ” button one time, LCD will show 

“  MX  ”  mark  to  indicate  the  recorded  Maximum 
value and maximum value occurred Elapsed Time.

4. Press “ MX/MN ” button one time, LCD will show 

“ MN ” mark to indicate the recorded Minimum value 
and minimum value occurred Elapsed Time.

5. Press “ MX/MN ” button one time, LCD will show 

“ MX/MN ” mark to indicate the current value and 
current Elapsed Time.

6. Step 3, 4, 5 can be cycled.
7. Press “ MX/MN ” button for 2 seconds to exit this 

mode.
Note : The Elapsed Time intial setting is “ minute : 

second  ”,  when  current  Elapsed  Time  is 
over 60 minutes, the Elapsed Time will be 
auto  setting  to  “  hour  :  minute  ”.  The 
maximum  Elapsed  Time  length  is  100 
hours.

0

4.  Function selector :

For selection of desired function.

5.  Range button :

A. In  ACA,  ACV,  DCV, 

m

A,  Capacitance  and 

Resistance function.
1. Press it once to enter the manual range select 

mode, LCD will show ® mark.

2. Press again to select the desired range.
3. Press it 2 seconds to exit the manual range and 

enter to the autorange mode, LCD ® mark will be 
disappered.

B. In KW/HP/KVA function

a. Press  RANGE  button  to  select  viewing  the 

KW/hp,  KW/PF, KW/KVAR, KVA/

u

  or  V/A  dual 

display.

B1.Press  HOLD  button  keep  the  measured  valued 

in memory.

B2.Press  RANGE  button  to  select  viewing  the 

KW/hp,  KW/PF,  KW/KVAR,  KVA/

u

  or  V/A  dual 

display.

B3. Press HOLD button to exit this mode.
C. In TEMP function

0

0

1. Press RANGE button to select desired  C or  F

temperature units.

6.  Temperature input jack :

Only type K thermocouple input is accepted.

7.  LCD Display :

4 digital LCD with indications for measurement values, 
unit symbols, decimal point, polarity, over range and 
low battery; etc.

8.  COM Jack :

Connect  black  test  lead  for  voltage,  Power, 
Capacitance,  Resistance,  Diode,  Countinuity  and 
Micro ampere measurement.

9.  “ V ” Jack :

Connect  red  test  lead  for  Voltage  and  Power 
measurement as a positive terminal.z