Thank you for choosing this Mecmesin
instrument. With correct use it will give
many years of accurate and reliable
service.
Upon receiving the unit please check
that no physical damage has occurred
to the packaging material or the gauge
itself. If any damage is evident please
notify Mecmesin immediately.
When the low battery symbol appears
on the display, four new 1.5V AA
batteries should be fitted. Rechargeable
1.2V batteries should not be fitted to the
gauge. To release the battery cover at
the rear of the gauge press the battery
cover as indicated by the arrow and
slide the cover upwards. Ensure all
batteries are inserted the correct way
round. Incorrect insertion of batteries
can cause damage to the gauge. To
replace the battery cover slide it in from
the top edge of the gauge until it clicks
into place.
If the gauge is not going to be used for
3 months or more the batteries should
be removed and stored separately.
(Settings and calibration data will
not
be lost if batteries are removed).
The gauge can be powered on by
pressing the large ‘ON/ZERO’ key.
After a short self test during which the
software version number and revision
date are shown, the gauge then
displays the transducers range before
entering the main display mode.
After the self test and providing no load
has been applied to the gauge, the
display will show zero, this is because
the gauge re-zeros itself as part of the
self test routine.
Fitting and
Replacing
Batteries
Powering Up
More than 120% of full scale load is
being applied to the transducer
‘-0L-’
Physical
overload
‘0L’
Maximum
reading overload
(not physical)
‘t-Err’
Tare error
‘No Ao’
No Auto
off
‘C-dEF’
Calibration default
Operating Instructions
Display Messages
Before Use
Automatic power off is disabled
The zero function was performed
while the transducer was in an
overloaded state
The peak force reading was an
overload
DEALER STAMP
431-
051
-
02
Operating Manual
For a copy of the EC Declaration of Conformity please refer to:
www.mecmesin.com/literature/cedeclarations.html
Declaration of conformity supplied as standard, Calibration
certificate available on request
Force and Torque Test Solutions
Mecmesin Limited
Newton House
Spring Copse Business Park
Slinfold
West Sussex
RH13 0SZ
England
T +44 (0) 1403 799979
F +44 (0) 1403 799975
E [email protected]
www.mecmesin.com
Force
Gauge
Compact
Compact
Calibration data has become invalid
please notify Mecmesin or an authorised
representative for re-calibration