Mechatron PHP-10 User Manual Download Page 5

 

 

 

 

2.  

The connections to the hand held test system are sealed with ‘O’ ring, bonded seals 
& Nylon washer to avoid leak. 

Ensure that the adapters are tightened sufficiently. 

Allow  the  pressure  to  settle  for  1-2  minutes  before  taking  the  reading  due  to 
thermodynamic effect, setting of seals and expansion of flexible hose. 

 

3.

  

 

Red line indicator   

 

                           The Fine adjustment should not be forced to move beyond the red 

 

 line indicator on the body 

 

2.2 TECHNICAL SPECIFICATIONS 

Pressure Medium 

Air 

Working Pressure 

-0.90 to 10/20/25/35/40/60 

Vernier Adjuster 

Micro-adjustable, Volume control 

Minimum resolution of Vernier 

1 mbar 

Test connection port 

1/4” BSP (F) 

Gauge/ Indicator connection port 

3/8” BSP (F) 

Weight 

750 grams (appox.) 

Stabilization time 

– 2 minutes 

Dimension 

230 x 65 x 130 mm (L x B x W) 

Summary of Contents for PHP-10

Page 1: ...AND PUMP MECHATRON CALIBRATION AND INSTRUMENT 19 37 2ND Main Road Sabari Nagar Mughalivakkam Chennai 600116 Tamilnadu India Phone 9940060420 9962060420 Mail info mechatroncalibration com Website www m...

Page 2: ...PNEUMATIC HAND PUMP PHP 10 20 25 35 40 60...

Page 3: ...factured with high quality for calibrating analog gauges by comparison measurements It can be used for lab and field applications There is a pressure vacuum changeover switch in the pump which is used...

Page 4: ...d intervention and or use FEATURES Portable Light in weight Low Cost Rugged Construction SAFETY INSTRUCTIONS 1 Check for system fittings and connection tight and leak free Tighten the pressure release...

Page 5: ...expansion of flexible hose 3 Red line indicator The Fine adjustment should not be forced to move beyond the red line indicator on the body 2 2 TECHNICAL SPECIFICATIONS Pressure Medium Air Working Pres...

Page 6: ...PUMP ASSEMBLY...

Page 7: ...F Swivel adaptor 7 Nylon washer 8 Plastic plug 9 Flexible hose to item under test 10 Pressure port 3 8 BSP Female connection 11 Nylon washer 3 8 ACCESSORIES STANDARD 1 Hose 1 meter with swivel adaptor...

Page 8: ...To adjust the maximum output pressure of the system turns the knurling thumb nuts 2 to increase or decrease the stroke length so that the pressure is controlled 4 0 OPERATING PROCEDURE Connect the mas...

Page 9: ...adjustment valve 4 for a precise pressure vacuum adjustment After the calibration completely release the pressure vacuum by twist off the vent valve Release valve gradually in anticlockwise direction...

Reviews: