Mechatron PHP-10 User Manual Download Page 3

 

 

 

Symbol Used 

Sl. No. 

Symbol 

Description 

 

Read the user manual before operating the 
instrument. 

 

Warning 

– Condition that may pose hazards 

to the user 

 

Caution 

– Condition that may damage the 

instruments 

 

Special Information 

 

INTRODUCTION: 

PHP-10/20/25/35/40/60 bar hand pump is a pneumatic pressure / vacuum generating 

unit which are specially designed for testing and calibrating pressure instruments. This 

pump is  manufactured  with  high  quality for calibrating  analog  gauges  by comparison 

measurements. It can be used for lab and field applications. 

There  is  a  pressure  /  vacuum  changeover  switch  in  the  pump,  which  is  used  for 

selecting either vacuum or pressure output. It is fitted with a fine adjustment valve and 

release  valve.  The  output  pressure  /  vacuum  of  the  hand  pump  can  be  adjusted 

precisely  by  the  fine  adjustment  valve  while  calibration.  Release  valve  is  used  for 

releasing the pressure / vacuum after calibration. The output pressure /vacuum of the 

pump is in the range of -0.90 to 10/20/25/35/40/60 bar depending upon the model. 

Summary of Contents for PHP-10

Page 1: ...AND PUMP MECHATRON CALIBRATION AND INSTRUMENT 19 37 2ND Main Road Sabari Nagar Mughalivakkam Chennai 600116 Tamilnadu India Phone 9940060420 9962060420 Mail info mechatroncalibration com Website www m...

Page 2: ...PNEUMATIC HAND PUMP PHP 10 20 25 35 40 60...

Page 3: ...factured with high quality for calibrating analog gauges by comparison measurements It can be used for lab and field applications There is a pressure vacuum changeover switch in the pump which is used...

Page 4: ...d intervention and or use FEATURES Portable Light in weight Low Cost Rugged Construction SAFETY INSTRUCTIONS 1 Check for system fittings and connection tight and leak free Tighten the pressure release...

Page 5: ...expansion of flexible hose 3 Red line indicator The Fine adjustment should not be forced to move beyond the red line indicator on the body 2 2 TECHNICAL SPECIFICATIONS Pressure Medium Air Working Pres...

Page 6: ...PUMP ASSEMBLY...

Page 7: ...F Swivel adaptor 7 Nylon washer 8 Plastic plug 9 Flexible hose to item under test 10 Pressure port 3 8 BSP Female connection 11 Nylon washer 3 8 ACCESSORIES STANDARD 1 Hose 1 meter with swivel adaptor...

Page 8: ...To adjust the maximum output pressure of the system turns the knurling thumb nuts 2 to increase or decrease the stroke length so that the pressure is controlled 4 0 OPERATING PROCEDURE Connect the mas...

Page 9: ...adjustment valve 4 for a precise pressure vacuum adjustment After the calibration completely release the pressure vacuum by twist off the vent valve Release valve gradually in anticlockwise direction...