EHC-PF
.
11
.
Press ENTER key to confirm the input and finish the setting, or use ON/OFF key to exit this mode.
NOTE: If the “Price” and “Item code” can’t be filled into the format you selected, “
OUT.P.R
” will show in weight
window by using PRINT key to communicate to other equipment unless “
232.out
” is set to “
HOST
”.
5.4.7
Then, the UNIT PRICE window shows “
USB.OUT
” (USB output), the TOTAL PRICE window shows “
none
” or
“
Host
”. Use the numerical keys 0, 1 to select enable or disable USB output if it installed in the scale:
0—
None
, disable USB output
1—
Host
, enable USB output, and it is connected with a host device, e.g. a PC
When the scale becomes stable, the data will be output from USB interface after pressing PRINT key.
5.4.8
Then, the WEIGHT window will show “
Setup
”, the UNIT PRICE window will show “
USB.bPS
” (vir tual RS232 band
rate: bit per second) and the TOTAL PRICE window shows baud rate xxxxx. Use numerical keys 1,2,3,4,5 to choose
RS232 baud rate:
1--1200bps
2--2400bps
3--4800bps
4--9600bps
5--19200bps
Press ENTER key for confirmation to go to the next step, or ON/OFF key to exit this mode.
5.4.9
Then, the UNIT PRICE window will display “
USB.DFT
” (vir tual RS232 data format), the TOTAL PRICE window will
display data format xxx. Use 1, 2, 3 key to select data format:
1—
8N1
8 bits data, no odd or even , 1 star t bit, 1stop bit
2—
7O1
7 bits data, 1 odd , 1 star t bit, 1stop bit
3—
7E1
7 bits data, 1 even, 1 star t bit, 1stop bit,
Press ENTER key to confirm the input and go to the next step, or use ON/OFF key to exit this mode.
5.4.10
Then, the UNIT PRICE window shows “
URT.CFT
” (communication format), the TOTAL PRICE window shows
communication format xxx. Use the numerical keys 0, 1, 2, 3, 4 to select the communication format:
0—
NONE
Non communication
1—
S-N
When the scale is stable, the data can be output after pressing PRINT key. That means the
current stable weight can be print out for many times if it keeps at stable status.
2—
S-1
When the scale is stable, the data can be output once after pressing PRINT key. That means
this stable weight can only be printed out for one time if it keeps at stable status. If one more
printing is need, the scale should become motion first, and then become stable again.
3—
Drs-1
After the scale is stable and PRINT key is pressed, the current data can be sent out one time
through serial communication por t. If another printing is need, the scale should be unloaded
first to make scale blow Min weight
1
, and then load the weight to make scale is over 20 e
1
4—
, at
last, to wait until it become stable.
Dsrs-1
When the scale is stable, press PRINT key to send the current data out for one time. If another
printing is need, the scale should be unloaded first to make scale blow Min weight, then wait
the scale to become stable, after that, load the weight to make scale is over Min weight, at last,
to wait scale to become stable again, after these, the new data can be sent out by pressing
PRINT key.
Press ENTER key to confirm the input and go to the next step, or use ON/OFF key to exit this mode.
NOTE: According the table, when item code or price is out of range, the printer will do nothing and the weight window
will display “
OUT.P.R
”.
5.4.11
RS-232 connects between scale and host: