INTRODUCTION

 

The MSI-6260CS Series crane scales are best calibrated using a remote indicator such as the MSI-
3750CS, MSI-9850, or the MSI-9750A Handheld. Calibration procedures using these devices are covered 
in their respective User Guides. If above listed remote devices are not in use with the MSI-6260CS, the 
only way to calibrate it is through the Comm. Port using the Terminal Access Mode. MSI recommends 
having test weights of at least 50% capacity to properly calibrate the Crane Scale. The calibration 
procedures will operate with as little as 12.5% of capacity, though rated accuracy cannot be guaranteed. 
For details of Multi-Point, Certified, or other specialized calibrations, refer to full version of MSI-6260CS 
User’s Guide. 

 
The following procedure describes how to perform a standard, single point 
calibration for a single load cell MSI-6260CS Crane Scale. Some menu items may 
appear slightly different than shown, depending upon actual software application 
version. However, those relative to Scale Calibration are accurate. 

 

REQUIRED EQUIPMENT: 

 

1) PC or terminal with terminal program and RS-232 serial port. 

 

2) Serial Communications cable, PN: 501705-0001. 

 

3) Known calibration test weight between12.5 and 100% of the scale’s capacity. 

 
 
ACCESSING EMBEDDED CELL SCALE 
 

The MSI-6260CS Crane Scale is controlled by an embedded MSI-9000 CellScale. All scale functions 
& operating parameters are stored within the MSI-9000. 

 

1)  Set the terminal program to emulate an ASCII terminal. 

 

2)  Set the terminal programs port protocol to 9600 Baud, no parity, 8 data bits and 1 stop bit. 

 

3)  Disable flow control in the terminal program. 

 

4). Turn scale 

ON

 
5).  Connect the MSI-6260CS comport to the PC serial port with the 501705-0001 serial cable. 

 

6).  From PC, press 

ENTER

. Main Cell Scale Manu will appear.  

 Select 

Channels/Calibration

 press 

1

 
 
 
 
 
 
 
 

 
 
 
 
 
 
 

---------- MSI 9000 - CellScale ------------------------ 
-(c) 2005 Measurement Systems International - 
---------------------------------------------------------------- 
13:48:06 

 

   01/01/08 

 

(1)  

Channels/Calibration 

(2)  

Communications 

(3)  

Display 

(4)  

Feature Locks/Passwords 

(5)  

Product IDs 

(6)  

Set Points 

(7)  

System 

(8)  

About... 

 
Enter choice: 

Summary of Contents for MSI-6260CS

Page 1: ...MSI6260cs RF Digital Crane Scale Calibration Guide ...

Page 2: ...fferent than shown depending upon actual software application version However those relative to Scale Calibration are accurate REQUIRED EQUIPMENT 1 PC or terminal with terminal program and RS 232 serial port 2 Serial Communications cable PN 501705 0001 3 Known calibration test weight between12 5 and 100 of the scale s capacity ACCESSING EMBEDDED CELL SCALE The MSI 6260CS Crane Scale is controlled ...

Page 3: ...o b Standard Mode Industry c 4 to 20mA d Temperature Sensor Installed No ESC Exit Choose Channel Parameters 1 Continue These settings are correct 2 Choose the channel to calibrate 1 1 3 Calibration timeout None ESC Exit Channel Calibration 1 Perform a completely new calibration 2 Re do the current calibration 3 Manually adjust current cal points 4 Reset all cal parameters to defaults 5 Change disp...

Page 4: ...gest division size suitable for your application 7 Unload the scale make sure it is stable and press Y The scale performs its zeroing step removing all offset This process takes several seconds The number presented at the bottom of the screen is direct A D counts This is for troubleshooting only This number will never be steady as it represents 24 bits If the deviation is less than 250 counts the ...

Page 5: ...applied to scale there will be a significant increase in the Raw A D Counts Actual amount will vary depending upon rated mV V output of load cell and percentage of capacity applied to scale Channel Calibration Calibrating Ch 1 1 named Precision in Kilograms 10000 x 2 Test weight 0 in Kilograms 0 Test weight 1 in Kilograms 10000 Press Y if this is correct Press N to change Press B to backup Press E...

Page 6: ...irmation of data stored Disconnect serial cable from scale and cycle power Channel Calibration Calibrating Ch 1 1 named Precision in Kilograms 10000 x 2 Test weight 0 in Kilograms 0 Test weight 1 in Kilograms 10000 Test Weight 2 in Kilograms 10000 Press Y when scale is loaded and steady Press N to change Press B to backup Press ESC to complete and store calibration MSI 9000 CellScale c 2005 Measur...

Page 7: ... display HyperTerminal Comes with Windows Start Programs Accessories Communications HyperTerminal Hyperterm exe Name connection CSCom1 any name is fine Under connect using Direct to Com1 DO NOT TYPE PHONE NUMBER IN Hit OK Port Settings Baud Rate 9600 Data 8 bits Parity none Stop 1 bit Flow Control Hardware Hit OK May need to press space bar to refresh the display When done save session as you name...

Page 8: ...GHING SYSTEMS COMPANY Measurement Systems International TM 14240 Interurban Avenue South Suite 200 Ɣ Seattle WA 98168 4661 Ɣ USA Phone 206 433 0199 Ɣ Fax 206 244 8470 www msiscales com 2012 Rice Lake Weighing Systems ...

Reviews: