Safety voltages
Connect only voltages that are within the limits specified in this table.
Table 20. Safety specifica ons
Parameter
Condi ons
Specifica on
Voltage limits
Channel-to-earth
ground
±30 V max, Measurement Category I
Isola on
Channel-to-
channel
None
Channel-to-earth
ground
None
Note 10: Measurement Category I is for measurements performed on circuits not directly connected to the
electrical distribu on system referred to as MAINS voltage. MAINS is a hazardous live electrical supply system
that powers equipment. This category is for measurements of voltages from specially protected secondary
circuits. Such voltage measurements include signal levels, special equipment, limited-energy parts of
equipment, circuits powered by regulated low-voltage sources, and electronics.
Cau on! Do not connect the WebDAQ 504 to signals or use for measurements within Measurement
Categories II, III, or IV.
Signal connector
Table 21. Screw terminal specifica ons
Parameter
Specifica on
Connector types
4 BNC connectors
6-posi on screw terminal for digital connec ons
Terminal wiring
14 to 30 AWG copper conductor wire with 5 to 6 mm (0.20 to 0.24 in.) of insula on
stripped from the end
Screw terminal
Table 22. Screw terminal pinout
Terminal
Terminal
#
Label
Use
#
Label
Use
1
GND
Digital ground
4
DIO2
Digital bit 2
2
DIO0
Digital bit 0
5
DIO3
Digital bit 3
3
DIO1
Digital bit 1
6
GND
Digital ground
BNC connectors
Table 23. BNC connector pinout
Terminal
Terminal
#
Label
Use
#
Label
Use
1
AI0+
Channel 0 HI
5
AI2+
Channel 2 HI
2
AI0–
Channel 0 LO
6
AI2–
Channel 2 LO
3
AI1+
Channel 1 HI
7
AI3+
Channel 3 HI
4
AI1–
Channel 1 LO
8
AI3–
Channel 3 LO
WebDAQ 504 User's Guide
page 27