Testing Pulse Output
To verify that the board can perform a pulse output operation, do the following:
1.
Connect a scope to counter/timer 0 on the DT3010 Series board. Refer to
example of how to connect a scope (a pulse output) to counter/timer 0.
Note:
The Quick DataAcq application works only with counter/timer 0.
2.
In the Quick DataAcq application, choose
Pulse Generator
from the
Control
menu.
3.
Select the appropriate DT3010 Series board from the Board list box.
4.
Select either
Continuous
to output a continuous pulse stream or
One Shot
to output one
pulse.
5.
Select either
Low-to-high
to output a rising-edge pulse (the high portion of the total pulse
output period is the active portion of the signal) or
High-to-low
to output a falling-edge
pulse (the low portion of the total pulse output period is the active portion of the signal).
6.
Enter a percentage or use the slider to select a percentage for the pulse width. The pulse
width determines the duty cycle of the pulse.
7.
Click
Start
to generate the pulse(s).
The application displays the results both in text and graphical form.
8.
Click
Stop
to stop a continuous pulse output operation. One-shot pulse output operations
stop automatically.
Summary of Contents for Data Translation DT3010 Series
Page 2: ...DT3010 Series UM 16866 V User s Manual Title Page ...
Page 5: ......
Page 15: ...About this Manual 14 ...
Page 16: ...15 1 Overview Features 16 Supported Software 18 Accessories 19 Getting Started Procedure 21 ...
Page 23: ...Chapter 1 22 ...
Page 24: ...Part 1 Getting Started ...
Page 25: ......
Page 37: ...Chapter 2 36 ...
Page 49: ...Chapter 3 48 ...
Page 91: ...Chapter 4 90 ...
Page 102: ...Part 2 Using Your Board ...
Page 103: ......
Page 147: ...Chapter 6 146 ...
Page 159: ...Chapter 7 158 ...
Page 181: ...Chapter 9 180 ...
Page 193: ...Appendix A 192 ...
Page 213: ...Appendix B 212 ...
Page 231: ...Index 230 ...