Operational Problems
Problem Possible
Cause Correction
Display values not stable (i.e.
floating, fluctuating, etc.)
Loose interface connection
Make sure that all cable
connectors are properly
seated. Verify that all
connections inside the junction
box are secure.
Bad load cell connection
Check and, if necessary,
repair all load cell connections.
No Speed Readout
Bad electrical connection
Check all electrical
connections in the system.
Incorrect magnetic pickup to
gear air gap.
Verify that the air gap is set to
0.020”
Magnetic pickup failure
Replace magnetic pickup.