X-Y OPERATION: When [10] set at EXT/X, The oscilloscope is used for X-Y
operation, at this moment Input [18] is Y-axis with the same sensitivity, Input [13] as
X-axis, [14] can be adjusted continuously within 0.2V/DIV~0.5V/DIV.

3.3.4 TRIGGER SOURCE

In Fig 3-1, [12] provides 3 sources for selection, INT trigger

EXT trigger input from

[13], LINE input from power source.

3.3.5 HORIZONTAL SYSTEM OPERATION

(1)

 

Sweep setup : turn the sweep switch to correct position according to signal frequency,
adjust POSITION [7] to show the whole waveform within available area. Adjust with
VAR [8] if necessarily, trimming ratio is 

2.5:1.

(2)

 

There are 3 kinds of Trigger sources: [11] 

AUTO

auto sweep, when triggering

signal is applied, Level [9] will adjust to proper position.  The screen showing steady
free run wave required frequency higher than 20Hz;  [11] 

NORM

while waiting

for sweep, no trace; when receiving input signal, the circuits was triggered to sweep
and show waveform . [11] 

TV” used to determine TV signal.  The signal is

negative.

(3)

 

SLOPE selection: Used to select whether the trigger signal crosses trigger level in a

  positive-going or a negative-going direction.

(4)

 

Level setup : Use to adjust signal sweep on a level as starting reference.

3.3.6 SIGNAL CONNECT

(1)

 

Probe operating : 
Use 10:1 to 1:1 alternating probes.  When using 10:1 probe,  input impedance is
10M

 16 pF. While if 1:1 is used for observing lower signal, input impedance is

1M

 30 pF.  At this stage, please consider of the affection of the probe upon the

detected circuits.

(2)

 

Probe Adjustment

   Before using , 10:1 probe must be adjusted correctly, see Point 4.1.2

4   Measurement

4.1 EXAMINATION AND ADJUSTMENT BEFORE MEASUREMENT

In order to be more accurate and prevent errors, the following observations should be
done before taking measurement.

4.1.1 TRACE ROTATION

Generally, horizontal trace on the screen should parallel with the horizontal line.  Due to
the earth’s magnetic field and other factors which causes horizontal trace leaning and
making errors, you must examine the followings before taking measurement:
(1)

 

Adjust components on the front panel to obtain a horizontal trace on the screen .

(2)

 

Adjust vertical position and make sweep baseline lied in horizontal line of vertical
center

(3)

 

Examine whether sweep baseline is parallel with horizontal line, if not , please
following point 3.3.2(3) to correct it .

4.1.2 PROBE COMPENSATION

Adjustment of probe is useful to compensate mistakes cause by difference of

oscilloscope input characteristic:

(1)

 

Follow step 3.3.2, setup front panel components, and obtain a sweep baseline.

(2)

 

Set VOLTS/DIV to 10mV/DIV.

Summary of Contents for CQ5010B

Page 1: ...CQ5010B SINGLE TRACE 10MHZ OSCILLOSCOPE USER S MANUAL CONTENTS 1 SAFETY PRECAUTION 2 SPECIFICATIONS 2 1 VERTICAL SYSTEM 2 2 TRIGGER SYSTEM...

Page 2: ...nsitivity up to 50V div Sweep rate at 0 1S DIV 0 1 S DIV on horizontal system The oscilloscope is easy to operate and highly reliable It is an ideal instrument for research production education and de...

Page 3: ...otten into this instrument first switch the power off remove the power cord and call for repair z Do not place this instrument on an unstable place such as on a shaky stand or on a slant The dropping...

Page 4: ...Range 05 V 2 Frequency 1kHz 2 2 6 CRT Display Area 8 10DIV 1DIV 6mm Accelerating Voltage 1200V Display Color Green 2 7 POWER SOURCE Voltage Range 110V 10 Frequency 60Hz 2Hz Power Consumption 25W 2 8...

Page 5: ...2 1 50 20 10 5 V mV VOLTS DIV POSITION POSITION VOLT VAR CAL CAL TIME VAR CAL CAL LEVEL _ PROBE ADJUST FOCUS INTENSITY POWER 10MHz HANDY OSCILLOSCOPE AUTO NORM TV INT LINE EXT Y INPUT 1M 30pF 1M _ X E...

Page 6: ...gger level in a negative going direction EXT X X Y switch 11 AUTO NORM TV AUTO a single trace shown on screen even no signal Automatically reverts to triggered sweep operation when adequate triggered...

Page 7: ...stment a Slightly turn 20 22 counter clockwise but not to unscrew b Insert a straight end screw driver to 21 and observe the wave move screw driver in order to let the waveform parallel with the horiz...

Page 8: ...level as starting reference 3 3 6 SIGNAL CONNECT 1 Probe operating Use 10 1 to 1 1 alternating probes When using 10 1 probe input impedance is 10M 16 pF While if 1 1 is used for observing lower signa...

Page 9: ...TS DIV and observe waveform let waveform display on the screen within 5 divisions and turn VAR clockwise to the calibration position 3 Adjust level to make waveform steady 4 Adjust sweep controls to s...

Page 10: ...LTS DIV let waveform show in the proper position on the screen turn VAR to the calibration position 6 Read the divisions between the zero reference level to the wave form by the tested object 7 Calcul...

Page 11: ...eform on the same horizontal level 5 Measure the distance between the two points the measurement is calculated by the following equation Distance between 2 points DIV rate of sweep t Div TIME S Horizo...

Page 12: ...o coincide with a vertical reference line 6 Measure the horizontal distance in divisions between the 10 and 90 points on the waveform divisions 7 Pulse rise time and fall time measurement is calculate...

Page 13: ...of sweep signals signals of Lissajous pattern or other equipment s display setup X Y mode operation is turn 10 to EXT X input signals through 13 sensitivity to be adjusted directly with 14 then input...

Reviews: