MCP CQ5010B User Manual Download Page 3

OS-10V                                                                                                                             02/04/2001 Rev.2

2

Non compliance with the warnings and/or the instructions for use may damage the
instrument and/or its components or injure the operator.

Take extreme care for the following conditions when using the instrument :

z

 

For your own safety and that of the instrument, you must follow the procedures described
in this instruction manual and especially read all the notes preceded by the symbol 
carefully

.

z

 

Do not use this instrument in a location where there is explosive gas in the vicinity.  The
use of this instrument in a location where there is explosive gas could result in explosion.

z

 

If there is any smoke, abnormal odor, or abnormal sound coming display type this
instrument, immediately switch off the power and disconnect the power cord.
Continuous using of this instrument under these conditions could result in electrical
shock or fire.  After disconnecting the power cord, contact the service offices for repair.
Repair by the user is dangerous and should be strictly avoided.

z

 

Take care not to allow water to get into this instrument or the wetting of the instrument.
The use of this instrument in a wet state could result in electrical shock or fire.  If water
or other foreign matter has gotten into this instrument, first switch the power off, remove
the power cord and call for repair.

z

 

Do not place this instrument on an unstable place such as on a shaky stand or on a slant.
The dropping or turning over of this instrument could result in electrical shock, injury or
fire.  If this instrument has dropped or its cover has been damaged, switch the power off,
remove the power cord and call for repair.

z

 

Do not allow any foreign matter such as metal or inflammable substance to get in from
the air hole.  The entrance of any foreign matter from the ventilation port, etc., could
result in fire, electrical shock, or power failure.

z

 

Use this instrument with the rated AC power supply.  Use of this instrument with a
voltage other than specified could result in electrical shock, fire or power failure.  The
usable power voltage range is marked on the rear panel.

z

 

Do not remove either the cover or panel.

z

 

Take sufficient care when measuring high voltages.

z

 

Do not modify this instrument.

z

 

Avoid use of any damaged cable or adapter.

2.

 

SPECIFICATIONS

2.1 VERTICAL SYSTEM

Sensitivity

5mV/DIV. ~5V/DIV 

±

3%

Trimming Ratio

2.5:1

Rise Time

35ns

Bandwidth(-3dB)

DC:0~10MHz AC:10Hz~10MHz

Input Impedance

1M

 

±

3%,30 pF 

±

5pF

Max. Input Voltage

400V pk

2.2 TRIGGER SYSTEM

Trigger Sensitivity

Int 1 div., Ext 0.3V

Ext. Trigger Input Impedance

1M

 30pF

Ext. Trigger Max. Input Voltage

400Vpk

Trigger Sources

Int, Line, Ext

Trigger Mode

Norm, AUTO, TV

Summary of Contents for CQ5010B

Page 1: ...CQ5010B SINGLE TRACE 10MHZ OSCILLOSCOPE USER S MANUAL CONTENTS 1 SAFETY PRECAUTION 2 SPECIFICATIONS 2 1 VERTICAL SYSTEM 2 2 TRIGGER SYSTEM...

Page 2: ...nsitivity up to 50V div Sweep rate at 0 1S DIV 0 1 S DIV on horizontal system The oscilloscope is easy to operate and highly reliable It is an ideal instrument for research production education and de...

Page 3: ...otten into this instrument first switch the power off remove the power cord and call for repair z Do not place this instrument on an unstable place such as on a shaky stand or on a slant The dropping...

Page 4: ...Range 05 V 2 Frequency 1kHz 2 2 6 CRT Display Area 8 10DIV 1DIV 6mm Accelerating Voltage 1200V Display Color Green 2 7 POWER SOURCE Voltage Range 110V 10 Frequency 60Hz 2Hz Power Consumption 25W 2 8...

Page 5: ...2 1 50 20 10 5 V mV VOLTS DIV POSITION POSITION VOLT VAR CAL CAL TIME VAR CAL CAL LEVEL _ PROBE ADJUST FOCUS INTENSITY POWER 10MHz HANDY OSCILLOSCOPE AUTO NORM TV INT LINE EXT Y INPUT 1M 30pF 1M _ X E...

Page 6: ...gger level in a negative going direction EXT X X Y switch 11 AUTO NORM TV AUTO a single trace shown on screen even no signal Automatically reverts to triggered sweep operation when adequate triggered...

Page 7: ...stment a Slightly turn 20 22 counter clockwise but not to unscrew b Insert a straight end screw driver to 21 and observe the wave move screw driver in order to let the waveform parallel with the horiz...

Page 8: ...level as starting reference 3 3 6 SIGNAL CONNECT 1 Probe operating Use 10 1 to 1 1 alternating probes When using 10 1 probe input impedance is 10M 16 pF While if 1 1 is used for observing lower signa...

Page 9: ...TS DIV and observe waveform let waveform display on the screen within 5 divisions and turn VAR clockwise to the calibration position 3 Adjust level to make waveform steady 4 Adjust sweep controls to s...

Page 10: ...LTS DIV let waveform show in the proper position on the screen turn VAR to the calibration position 6 Read the divisions between the zero reference level to the wave form by the tested object 7 Calcul...

Page 11: ...eform on the same horizontal level 5 Measure the distance between the two points the measurement is calculated by the following equation Distance between 2 points DIV rate of sweep t Div TIME S Horizo...

Page 12: ...o coincide with a vertical reference line 6 Measure the horizontal distance in divisions between the 10 and 90 points on the waveform divisions 7 Pulse rise time and fall time measurement is calculate...

Page 13: ...of sweep signals signals of Lissajous pattern or other equipment s display setup X Y mode operation is turn 10 to EXT X input signals through 13 sensitivity to be adjusted directly with 14 then input...