MBW973-SF6_MANUAL_E_V3.1
31
Test Functions
Ice Test
The measuring accuracy can be checked with a simple, built-in test. The test may be performed
at any time, and is recommended if the results of your normal measurements do not correspond
to expectations and an error is assumed with the instrument.
Press the lower right menu key once to select the
Control
Setup
menu.
Control Setup
appears on the key while the
keys above change to the available menu options. Note that
the second key from the top indicates
Ice Test
.
Press the
Ice Test
menu key. A window requests you open
the measuring head.
The measuring head of the 973 is located on the right side of
the front panel. Remove the large tan cover screw by twist-
ing it counter-clockwise (about 3 full turns).
More detailed removal/installation information is available
beginning on page 42.