5.2
Battery
Check
When the life of the built-in battery comes to an end, the message
BAT
is indicated on the
display.
b
Whenever you open the display unit, check if
BAT
is indicated.
Replacing the battery
1
Undo the 8 screws at the rear of the display unit.
2
Lift the upper section of the enclosure on the left-hand side.
➝
The battery is located in the lower section.
3
Detach the battery cable from the electronics (pin-and-socket connector) and remove
the battery.
4
Check the enclosure seal.
b
If the seal is damaged:
replace the seal (spare parts
5
Insert the new battery and connect the battery cable.
6
Screw the enclosure back together. Make sure
– that the enclosure seal is correctly positioned
– that no cables are pinched.
7
Carry out a performance test (
) –
Caution:
for safety reasons, make this first
performance test after battery replacement outside hazardous locations.
HINWEIS:
Risk of damage by leaking battery
Caustic liquid could escape from the spent battery and could damage the
electronics of the display unit.
b
Remove a spent battery as soon as possible.
b
Replace the built-in battery at the latest after 3 years.
b
Note in a suitable place when the battery was last replaced.
WARNUNG: Risks in explosion-hazardous locations
b
When using the MLA 900 in hazardous locations, use only batteries of the
original type (
).
⊗
Never open the enclosure inside a hazardous location.