Radiation Leakage Testing
!
WARNING
To avoid risk of electrical shock, personal injury, or death, disconnect power to oven and discharge capacitor
before servicing, unless testing requires it.
16021520 Rev. 0
©2002 Maytag Appliances Company
16
WARNING
!
Check for radiation leakage after servicing. Should the
leakage be more than 4 mW/cm
2
inform Maytag
immediately. After repairing or replacing any radiation
safety device, keep a written record for future reference,
as required by D.H.H.S. and HEW regulations. This
requirement must be strictly observed. In addition, the
leakage reading must be recorded on the service repair
ticket while in the customer’s home.
Equipment
• Electromagnetic radiation monitor
• 600 cc glass beaker
Procedure for Measuring Radiation Leakage
Note before measuring:
• Do not exceed meter full scale deflection. Leak monitor
should initially be set to the highest scale.
• To prevent false readings the test probe should be held
by the grip portion of the handle only.
• The scan speed is equal to one inch per antenna
revolution or one inch per second if antenna speed is
unknown.
• Areas to be checked are all door seal areas and any
venting parts.
• Leakage with the outer panel removed ...4mW/cm
2
or
less.
• Leakage for fully assembled oven with door normally
closed ...4 mW/cm
2
or less.
• Leakage for a fully assembly oven (before the latch
switch (primary) is interrupted) while pulling the door
... 4 mW/cm
2
or less.
1. Pour 275 cc ±15 cc (9 oz ±1/2 oz) of 20°C ± 5°C
(68°F ± 9°F) water in a beaker which is graduated to
600 cc and place the beaker in the center of oven.
2. Set the radiation monitor to 2450 MHz and use it
following the manufacturer’s recommended test
procedure to assure correct results.
3. While measuring the leakage, always use the two inch
(5 cm) spacer supplied with the probe.
4. Press the start pad or turn on the timer and with the
magnetron oscillating, measure the leakage by holding
the probe perpendicular to the surface being
measured.
Measurement with the Outer Panel
Removed
WARNING
!
Avoid contacting any high voltage components.
Whenever you replace the magnetron, measure for
radiation leakage before the outer panel is installed
and after all necessary components are replaced or
adjusted. Special care should be taken in measuring
around the magnetron.
Measurement with a Fully Assembled
Oven
After all components including the outer panel are fully
assembled, measure for radiation leakage around the
door periphery, the door viewing window, the exhaust
opening, and air inlet openings.
Record Keeping and Notification After
Measurement
1. After any adjustment or repair to a microwave
oven, a leakage reading must be taken. Record
this leakage reading on the repair ticket even if it is
zero.
2. A copy of the repair ticket and the microwave
leakage reading should be kept by the repair
facility.
Summary of Contents for 1324704M
Page 4: ......
Page 40: ...2002 Maytag Appliances Company 16021520 Rev 0 B 1 Appendix B...