Maxtec Handi+ R218P12 Operating Manual And Instructions Download Page 4

      866.4.Maxtec    www.maxtec.com

II

 WARNINGS

 

!

Indicates a potentially hazardous situation, if not avoided, could result in death 
or serious injury.

»  Before use, all individuals who will be using this product must become thoroughly 

familiar with the information contained in this Operation Manual. Strict 

adherence to the operating instructions is necessary for safe and effective product 

performance. This product will perform only as designed and only if installed and 

operated in accordance with the manufacturer's operating instructions.

»  Although the sensor of this device has been tested with various anesthesia gases 

including nitrous oxide, Halothane, Isoflurane, Enflurane, Sevoflurane and Desflurane 
and found to have acceptably low interference, the device in entirety (including 
electronics) is not suitable for use in the presence of a flammable anesthetic mixture 
with air or with oxygen or nitrous oxide. Only the threaded sensor face, flow diverter, 
and “T” adapter may be allowed to contact such a gas mixture. 

» This device does not contain automatic barometric pressure compensation. 

  

DO NOT 

use near any type of flame or flammable/explosive substances, vapors 

 or atmosphere.

» Not for use in an MRI environment.

»  Improper use of this device can cause inaccurate oxygen readings which can lead to 

improper treatment, hypoxia or hyperoxia. Follow the procedures outlined in this user 
manual.

CAUTION:

 

Indicates a potentially hazardous situation, if not avoided, could 

result in minor or moderate injury and property damage.

»  The Maxtec MAX-250 oxygen sensor is a sealed device containing a weak acid 

electrolyte, lead (Pb), and lead acetate. Lead and lead acetate are hazardous waste 
constituents and should be disposed of properly, or returned to Maxtec for proper 
disposal or recovery. 

»   The  Handi+ is not intended for steam, ethylene oxide or radiation sterilization.   

 DO NOT 

autoclave or expose the sensor to high temperatures.

 DO NOT 

immerse the Handi+ oxygen analyzer in any cleaning solution.

»  The flow diverter provided with the Handi+ is for use with flowing gases only.

 DO NOT 

use the diverter when performing static sampling (e.g., in incubators,  

  oxygen tents, oxygen hoods).

 DO NOT 

attempt any repairs or procedures, which are not described in this   

 Operation Manual. Maxtec cannot warrant this product from damage resulting from  
 misuse, unauthorized repair or improper maintenance of this product.

  »  The Federal (USA) law restricts this device to sale by or on the order of a physician.

»  There are no internal user-serviceable parts.
»  Never install the sensor in a location that will expose the sensor to patient’s exhaled 

breath or secretions.

Summary of Contents for Handi+ R218P12

Page 1: ...H Handi O P E R A T I N G M A N U A L I N S T R U C T I O N S F O R U S E R 2 1 8 P 1 2 R218M12 Rev J CAL...

Page 2: ......

Page 3: ...s from the date of shipment from Maxtec under normal operating conditions and provided that the Handi is properly operated and maintained in accordance with Maxtec s operating instructions Should Hand...

Page 4: ...Improper use of this device can cause inaccurate oxygen readings which can lead to improper treatment hypoxia or hyperoxia Follow the procedures outlined in this user manual CAUTION Indicates a potent...

Page 5: ...ation A 3 2 4 Direct Flow Calibration Configuration B 4 3 0 OPERATING THE HANDI 4 4 0 FACTORS INFLUENCING ACCURATE READINGS 5 4 1 Elevation Changes 5 4 2 Temperature Effects 5 4 3 Pressure Effect 5 4...

Page 6: ......

Page 7: ...blank The analyzer turns off after 1 3 minutes if no buttons are pressed Calibration Button Press the calibration CAL button to adjust the calibration value to reflect the known oxygen concentration T...

Page 8: ...his oxygen analyzer ideal for portable oxygen analysis by qualified heath care professionals 2 0 CALIBRATING THE Handi 2 1 Before You Begin A protective film covering the threaded sensor face must be...

Page 9: ...encing Calibration and Performance on page 5 of this manual Note We recommend use of medical grade USP or 99 purity oxygen when calibrating the Handi 2 3 In Line Calibration Configuration A 1 Put the...

Page 10: ...ess the CAL button on the Handi The calibration gas value on the analyzer display should read 20 9 or 100 depending on the gas stream used 3 0 OPERATING THE HANDI To Check the Oxygen Concentration of...

Page 11: ...e Handi will hold calibration and read correctly within 3 when in thermal equilibrium within the operating temperature range The device must be thermally stable when calibrated and allowed to thermall...

Page 12: ...an error code occurs E03 No valid calibration data available Make sure unit has reached thermal equilibrium Press and hold the Calibration Button for three seconds to manually force a new calibration...

Page 13: ...point on display located after the first digit Accuracy and Linearity 1 of full scale at constant temperature R H and pressure when calibrated at full scale Total Accuracy 3 actual oxygen level over...

Page 14: ...on Dioxide 10 2 Helium 70 2 8 0 HANDI SPARE PARTS AND ACCESSORIES 8 1 Standard Replacement Parts and Accessories Part Number Item R218M12 Operation Manual Handi R218M12 001 Operation Manual Handi Inte...

Page 15: ...calibration value Make sure that the calibration gas is connected to the Handi at 2 liters per minute for a minimum of 2 minutes prior to proceeding with calibration This 2 minute equilibration time...

Reviews: