CONTROLLER MENUS: 1Y.5 HARDWARE IOS
7 - 20
D-MAXE with OI-B/OI-N
www.maxcessintl.com
MI 2-296 1 D
Menus 1y.5.1.1
S 01 (X5 /1)
These menus are used to calibrate sensor S01 at input X5/1.
Menus 1y.5.1.2
S 02 (X5 /2)
These menus are used to calibrate sensor S02 at input X5/2.
Menus 1y.5.1.3
S 03 (X9 /1)
These menus are used to calibrate sensor S03 at input X9/1.
(On a D-MAXE 3, X9/1 is also connected to X11/2.)
Menus 1y.5.1.4
S 04 (X9 /2)
These menus are used to calibrate sensor S04 at input X9/2.
(On a D-MAXE 3, X9/2 is also connected to X11/1 and is
referred to as X11/1 in the menus.)
General information about
these menus
The analog sensors present in the system must be calibrated.
Each sensor in the system must be calibrated on the controller
to which the sensor is connected.
Note:
It is possible that special sensors are used in the customer's
system, for example a capacitive sensor or a camera. Sensors of
this type cannot be calibrated with the procedure described
here. The information required for a calibration is available in a
set of special operating instructions for the sensor or in
"Supplementary Operating Instructions“ in the system
documentation.
The following sections describe the
1y.5.1._.1 Calibration
1y.5.1._.3 Dimension
1y.5.1._.4 Sensor Supervision menus.
These explanations apply to all four analog sensor connections.
The S1 sensor on the X5 /1 connection is used as an example
for the explanations.