Moving average 
[times] 

:  Output  Ave.  value  of  past  output  values  among  the  specified  times  on  the 

parameter for each PM counting. If it is “1”, moving average is not performed. 
Initial value 1 [times]    Range of setting 1 to 60 [times] 

Cleanness conversion 
selection 

:  Select target to calculate conversion value of cleaning level on the list box. 

Initial value “PM10 count” 
Options “Each PM Mass concentration/Each PM Counting” 

Cleanness conversion 
coefficient 

:  Specify factor to calculate conversion value of cleaning level. 

Conversion value of cleaning level = Value of target x Conversion factor 
Initial value 1000    Range of setting 1 to 1,000,000 

(#1) If switching the selection of range, max. range, On threshold and Off threshold of the 

specified selection of range will return to initial value. 

(#2) In case of lowest value setting of measurment value, “4mA” will be output. 

(#3) Regardless of specified values, display of measurement result is available up to “9999.999 

µg/m

3 “

(#4) Please specify a value of On threshold bigger than a value of Off threshold. 
        And, if specifying Of threshold = max. range, contact point will never be ON (always OFF). 
        Please specify On threshold = max. range, if using contact point output as abnormal output. 
# Updating of current output is 20 seconds on each mass concentration and the same value of 

accumulated time of counting per each counting. 

Conversion value of cleaning level will be changed by conversion value or counting like shown 
in the above. 

 

9-4. Screen of Logging 

This is the screen of setting of logging. 
Outputting CSV data of measuring results on the Air particle monitor is available. 
Press the “Apply” button to update after modified sampling cycle of logging and target of logging. 
Please refer to decide a maximum range and a threshold. 

 
 
 
 

 
 
 
 

 
 
 
 

 
 
 
 
 
 
 
 
 
 
 

(A) Logging 

Operation of logging 

“Start” button 

Start logging. 
During logging, the “Start” button” is disabled and the “Stop” button 
is enabled. 

(A) 

(B) 

Fig. 13: Screen of Logging setup 

Summary of Contents for PFM-AP01

Page 1: ... 7 Wi Fi communication 5 8 Wi Fi access start up 6 9 Web browser screen 10 10 Troubleshooting 17 OPERATION MANUAL FOR AIR PARTICLE MONITER TYPE PFM AP01 1 8 18 Norimatsu Higashi Yahatanishi ku Kitakyushu 807 0837 Japan Phone No 8193 691 3731 Fax No 8193 691 3735 http www matsushima m tech com Certainly perform the right operation by operators perusing this Manual ...

Page 2: ...after surely confirming the specification list Keep the installation environments of the temperature the operating voltage frequency etc Confirm setting wires before running Do not fall the product and make a hard impact to the product Product damages can be occurred Surely wire the necessary terminals ground terminal etc Make sure to take off all wires if operating electric welding near the produ...

Page 3: ...ased on the above class over 10 000 6 Output signal Analog output DC4 to 20mA Resolution 1µA Load resistance Max 500Ω Contact output 1c Contact point capacity DC30V 3A AC250V 3A 7 Display RGB full colored LED 1 1 8 Cumulative count time 1 to 60min In case of 1min Updating number of particle counting per 1min 9 Update time of mass concentration 20s 10 Moving average number of particle counting 1 to...

Page 4: ...rder on devices Green Orange Blue flashing Under mock output testing mode B Air inlet To intake floating particles in the air C Air outlet To exhaust floating particles D Wi Fi mode switch To switch Wi Fi communication mode E Reset button To initialize SSID and password If you hold the reset button for 3 seconds the SSID and password are reset to their default values F Terminal block Leg spring te...

Page 5: ...arricade Please do mounting with confirming an available area for Wi Fi communication If any barricade or stepping over rooms and floors connection will be not available or to be unstable 5 2 Mounting Please do mounting Product using holes for mounting M3 screw holes at 2 positions Please refer the recommended mounting directions on Fig 2 Recommended mounting directions Mounting directions not to ...

Page 6: ...ne PC or one tablet Not available to access more than one terminal to a Product in the same time 7 2 Station mode ST This is the mode using a wireless LAN router measuring display and modifying on Product can be available on a PC or a tablet connected to the wireless LAN of Product If ST accessing to Product please access from only one terminal In case of accessing from more than one terminal comm...

Page 7: ... network Please surely select Yes If select No unable to access to the network In case of AndroidOS iOS a Please access to the network PFM AP01 from your individual terminal Tablet Mobile phone Displaying last triple digits of the MAC address on the product tag for individual identification Input the password matsushima to access to the network Please tap on the notice of restricted network access...

Page 8: ...will be blue light Wi Fi access to the device can be completed Changing the status LED may take over 20 seconds Select Fig 7 LED Color Tap Fig 6 2 Wi Fi Access In case of AndroidOS iOS matsushima Input Select Wi Fi access ...

Page 9: ...vating the web browser screen the password login screen will be displayed Please input matsuap and press the button OK If wrong inputting Authorization Required is displayed on the screen Please input 10 10 10 1 on the address bar of web browser again and re enter the password After corrected the password Please select English on the Language Setting Display of Japanese will be changed to display ...

Page 10: ...g the network setting the Wi Fi module will be restarted and Wi Fi access is released once 8 2 Method of connection after the initial access start up Access point mode If modified SSID and the password of Access point mode and Wi Fi access was released once so please perform Wi Fi access again according the above Station mode In case of using Station mode please switch on the Wi Fi mode SP Station...

Page 11: ...oving to each item will be available A Product logo Move to top screen from each item from B Measurement result Display measuring result of Product C Measurement setting Set up parameters for measurement D Logging Specify parameters for logging E Network setting Specify parameters for Wi Fi access F Output setting Perform mock output testing of current and contact point G Calendar Display present ...

Page 12: ...e is each 20 seconds Displaying range is 0 000 to 9999 999 µg m3 PM2 5 mass concentration PM10 mass concentration B PM1 0 Count Display floating particle counting cnt detected within accumulated time of counting Updating cycle is specified by accumulated time of counting with calculation Initial value of accumulated time of counting is 1 min it means updating each 1 minute initially Displaying ran...

Page 13: ...µg m3 3 Count of each PM Initial value 5 000 cnt Range of setting 1 to 1 000 000 cnt Cleanness conversion value Initial value class 100 000 Range of setting class 1 to 1 000 000 B Contact output setting Specify threshold of contact point output ON or OFF 4 On threshold Value to be ON for contact point output when each measuring value is bigger Contact point will be ON if bigger measuring value tha...

Page 14: ...egardless of specified values display of measurement result is available up to 9999 999 µg m3 4 Please specify a value of On threshold bigger than a value of Off threshold And if specifying Of threshold max range contact point will never be ON always OFF Please specify On threshold max range if using contact point output as abnormal output Updating of current output is 20 seconds on each mass conc...

Page 15: ...pling time when the start button is pressed For example the first logging data will be recorded in one minute later than when the start button is pressed if a cycle of updating is set as 1 min Log data is saved on the connected Wi Fi terminal So logging is not available if shutting the web browser during logging disconnecting sleep mode and screen lock on the OS and background action AndroidOS iOS...

Page 16: ...haracters by one byte Min number 1 character Password Initial value matsushima Max 32 alphanumeric characters by one byte Min number 8 characters DHCP Initial value Enable Range Select Enable or Disable 1 IP address Initial value 10 10 10 1 Range 0 to 255 x 4 Subnet mask Initial value 255 255 255 0 Range 0 to 255 x 4 B Station mode Mode through wireless LAN router between Product and PC Tablet Dis...

Page 17: ...the selected current value Press the END button to return to normal output C False alarm setting Specify actions under trouble on Product If modifying the setting please press the Apply button to update Contact output Select ON or OFF Initial value ON Current output Select among 3 6mA 4mA 20mA 22mA HOLD Initial value 4mA D Auto cancellation Select Automatic release of output testing After modified...

Page 18: ...utput test mode to be completed If not completed please press the End button to be completed If completed please set power OFF ON 6 On work on contact point 7 No changing on current output value 8 No display of result on the screen of measurement result Please update the web browser Please try to check the terminals to be confirmed and browser as described in Section 8 1 c 9 All measuring results ...

Reviews: