![Matsushima GRAPHIC COM4 Instruction Manual Download Page 28](http://html1.mh-extra.com/html/matsushima/graphic-com4/graphic-com4_instruction-manual_1741925028.webp)
27
9. Current output test
Outputs entered simulation value in order to test the functioning of the current output.
(Level input range: -10.00
%
to +110.00%,
current input range: 3.55mA(3.60mA
:
Microwave Level Meter under Ver2.0) to 22.00mA)
Important
:
When you wish to end the simulation and return the instrument to actual
measurement, please push 'Esc' button.
12.345m
Distance
Menu
0
1. Identification setting
2. Basic setting
3. Sensor adjustment
4. Graph display
5. Self test
6. Reset
▼
Menu
0
2. Basic setting
3. Sensor adjustment
4. Graph display
5. Self test
6. Reset
7. Current output test
▼
Menu view will be
displayed.
Push
『+』
six times
to shift down to
Current output test.
Push
『Ent』
to
select Current
output test.
Current output test 7
Percent = 100.00%
Current = 20.000mA
OK
Output = %
= mA
Current ouput test
view will be
diaplayed.
Example: Do test for
50.000%.
Push
『
Ent
』
, then
value to be changed
will be indicated by
block cursor.
Current output test 7
Percent =
1
00.00%
Current = 20.000mA
OK
Output = %
= mA
Use
『+』
or
『
®
』
and change 100.0%
to 50.0%.
Push
『Ent』
set to
entered value.
Current output test 7
Percent = 50.00%
Current = 12.000mA
OK
Output = %
= mA
Once percent value
setcurrent value will be
automatically calculated
and displayed.
Push
『+』
twice and shift
cursor to OK.
+
× several times
→
or
Current output test 7
Percent = 50.00%
Current = 12.000mA
OK
___
Testing
_
Output = %
= mA
Push
『
Ent
』
to start
simulation.
When simulation is active
message
「
Testing
」
will
be displayed. Also active
test values are displayed
below the message.
Current output test 7
Percent = 50.00%
Current = 12.000mA
OK
__ _
Testing
_
Output = 50.00 %
= 12.00 mA
Push
『
Esc
』
to interrupt
test.
You can change percent
or current value and
continue the test.
When
test
completed
push
『
Esc
』
three times
Measurement view
Ent
× 1 click
Ent
× 1 click
Ent
× 1 click
Ent
× 1 click
Ent
× 1 click
+
× 6 click
+
× 2 click
Esc
× 3 click
【
Measurement view
】
Push
『Ent』
to enter