Setting Functions   

KAS/KAS-R series, -LRs/-LRs4 option 

3

 

Setting Functions 

 

3-1

 

Communication Protocol 

Baud rate 

Asynchronous 9,600 [bps] (Fixed) 

Character length 

8 [bits] 

Stop bit 

1 [bit] 

Parity 

None 

Flow control 

None 

The delimiter is CR (0D

H

) and fixed. 

 

For -LRs4 option, it will be a two-wire transmission (half-duplex). 

 

3-2

 

Communication Option Connector Pin Assignment 

 

3-2-1

 

-LRs option 

(1)  TxD (Transmission) 

(5)  NC 

(2)  RxD (Reception) 

(6)  NC 

(3)  GND 

(7)  GND 

(4)  NC 

(8)  GND 

 

3-2-2

 

-LRs4 option 

(1)  TxD (+)/RxD (+) 

(5)  TxD (+)/RxD (+) 

(2)  TxD (-)/RxD (-) 

(6)  TxD (-)/RxD (-) 

(3)  GND 

(7)  GND 

(4)  Termination resistor ON/OFF 

(8)  GND 

 

3-2-3

 

Termination Resistor (for -LRs4 option) 

 

A  termination  resistor  for  RS-485  is  available  by  shorting  between  pin  (4)  and  pin  (8)  of  the 
communication option connector. When they are not shorted, there is no termination resistor. 
Set the termination resistor if necessary. 
 
Example of termination resistor setting: 
Set the termination resistor to the first and end connection devices of the RS-485 connection. 
Do not set it to any other devices. 

 
 
 
 
 
 
 
 
 

(1)    (2)      (3)      (4) 

(5)    (6)      (7)    (8) 

TxD (+)/RxD (+) 

TxD (-)/RxD (-) 

Device 

Device 

Device 

Device 

Termination resistor: Presence 

Termination resistor: Presence 

Termination resistor: Absence 

Do not connect anything to NC. 

Summary of Contents for KAS Series

Page 1: ...Instruction Manual F RA 001 3R3 MODEL KAS KAS R series LRs option LRs4 option B N 199 9 011 199 9 011 Rev 0 1 ...

Page 2: ...ecause Matsusada Precision is constantly striving to improve its high quality products the information contained in this manual is subject to change without notice Every precaution has been taken in the preparation of this manual Nevertheless Matsusada Precision assumes no responsibility for errors or omissions Neither is any liability assumed for damages resulting from the use of the information ...

Page 3: ... 1 LRs option 3 3 2 2 LRs4 option 3 3 2 3 Termination Resistor for LRs4 option 3 3 3 Connection Condition for Communication Option Connector 4 3 4 Setting Unit Number 4 4 Digital Operation Check 5 4 1 LRs LRs4 Option 5 5 Matsusada Precision Original Commands 6 5 1 Command Format 6 5 2 Multi Command 6 5 3 Command Response String 7 5 4 Characters used for Commands 7 5 5 Command List 8 5 6 Command De...

Page 4: ......

Page 5: ... KAS KAS R series LRs LRs4 option 1 Overview 1 1 Overview 1 Setting output voltage 2 Setting output ON OFF 3 Measuring output voltage current 1 2 General Information Accessories Communication option connector 1 pc ...

Page 6: ...2 External View 2 KAS KAS R series LRs LRs4 option 2 External View 2 1 Rear Panel Communication option connector ...

Page 7: ... 2 TxD RxD 6 TxD RxD 3 GND 7 GND 4 Termination resistor ON OFF 8 GND 3 2 3 Termination Resistor for LRs4 option A termination resistor for RS 485 is available by shorting between pin 4 and pin 8 of the communication option connector When they are not shorted there is no termination resistor Set the termination resistor if necessary Example of termination resistor setting Set the termination resist...

Page 8: ... 4 ST 3 5 1790124 for printed circuit board produced by PHOENIX CONTACT In order to connect the communication option connector use a cable compatible with the following terminal connection standards Input cable AWG24 to 16 Stripping length 10mm 3 4 Setting Unit Number It is set to 0 at the factory shipment The unit number is changeable with a command Although it can be selected from 0 to 31 make s...

Page 9: ...it and turn on the power 3 Set the unit number Transmit AL UNIT 0 from the RS 232C RS 485 controller If it is in a default state this operation is unnecessary because the unit number is set to 0 4 Enable remote control Transmit 0 REN from the RS 232C RS 485 controller 5 Set an output voltage value Transmit 0 VSET xxx from the RS 232C RS 485 controller About the parameter xxx refer to 5 6 Command D...

Page 10: ...red 4 Set the unit number RS 232C RS 485 0 to 31 or AL A command with 20 or more characters including the delimiter is not accepted If lots of setting commands are transmitted in series commands may not be accepted due to the receive buffer overflowed In that case add a wait between each command 5 2 Multi Command When AL is specified as a unit number the command is applied to all connected units F...

Page 11: ... part in the ASCII code table Use LF 0AH and CR 0DH only as a delimiter It is case insensitive in alphabets used in commands Use a space 20H as a separator between unit number command and parameter in a command string Table ASCII code 1 SP C O SP L O CR 23H 31H 20H 43H 4FH 20H 4CH 4FH 0DH 0H 1H 2H 3H 4H 5H 6H 7H 0H NUL DLE SP 20 P p 1H SOH DC1 1 A Q a q 2H STX DC2 2 B R b r 3H ETX DC3 3 C S c s 4H...

Page 12: ...age value voltage value mode 9 Output OUT Sets ON OFF of output 9 Local Remote REN GTL Sets remote mode Sets local mode 9 9 Unit UNIT Sets Unit Number 9 Polarity POL Sets polarity 10 Measurement Output value Voltage VGET Measures output voltage voltage value mode 10 Current IGET Measures the output current value current value mode 10 Status check STS Checks the unit status 10 Only for KAS R series...

Page 13: ... OFF ON Format ex OUT ON OUT OFF Initial value OUT OFF Query In response to OUT it returns OUT OFF or OUT ON Remarks This supports Multi command REN command Enable remote control Format REN Initial value GTL Remarks When this command is received all commands will become available Receiving this command will turn on the REMOTE LED This supports Multi command GTL command Disable remote control Forma...

Page 14: ...tor the output current value Range 0 to the rated current value Resolution Refer to 5 7 Resolution of Command Setting and Return Values Unit mA Format IGET Response ex IGET 12 34 12 34mA is output STS command The unit status is returned Format ex STS Response unit number status Response ex 1 CO LO Unit 1 is in a state of output enabled and local mode 1 CF RM Unit 1 is in a state of output disabled...

Page 15: ...han 2000kV 0 1kV 20kV or more to less than 200kV 0 01kV 2 kV or more to less than 20kV 0 001kV less than 2kV 0 0001kV IGET 2000mA or more 1mA 200mA or more to less than 2000mA 0 1mA 20mA or more to less than 200mA 0 01mA 2mA or more to less than 20mA 0 001mA less than 2mA 0 0001mA Revision History Rev No Rev Date Revision Contents 0 0 2020 12 First edition 0 1 2021 01 Added KAS R series modified t...

Page 16: ......

Reviews: